(15) Results were found. Showing 1-10.
What is the Data Format of the ASCII File for the DL2700?
The data is saved in blocks of 8-bits and is expressed in decimal notation. In other words, if all 8-bits are zeroes, the decimal value is 0; if all 8-bits are 1s, the decimal value is 255....
item: 1218 | http://tmi.yokogawa.com/ar/service-support/faqs/1218/
How Long Does it Take to Transfer Measured Data from the DL2700 to the PC via GPIB?
The data transfer rates for binary data is as follows: 1 kW 0.035s 10 kW 0.180s 50 kW 0.779s 100 kW 1.536s 250 kW 3.812s 500 kW 7.527s 1 MW 15.067s 2 MW 30.000s The...
item: 1227 | http://tmi.yokogawa.com/ar/service-support/faqs/1227/
What is the Dead Time During Sequential Store on the DL2700?
The dead time is approximately 10µs....
item: 1221 | http://tmi.yokogawa.com/ar/service-support/faqs/1221/
Why is it That Only 30 Files can be Saved to the Floppy Disk or External MO and Error:Code 612 is Displayed if Attempting to Save Additional Files on the DL2700?
When saving measured data or set-up data, there is a limitation on the number of files that can be saved depending on the data format. The number of files that can be saved for the data format is listed below: Waveform ...
item: 1217 | http://tmi.yokogawa.com/ar/service-support/faqs/1217/
What is the Service Procedure for Verifying Time Axis Accuracy?
The Time Axis Accuracy is the accuracy of the A/D clock. It is not possible to directly measure the A/D clock and the A/D clock is not output from the DL850/SL1400. To verify the accuracy of the A/D clock, you will nee...
item: 1288 | http://tmi.yokogawa.com/ar/service-support/faqs/1288/
What is Meant by the "Setup Time (15 ns Typical) With Respect to the Edge of the Logic Trigger?"
The following statement "setup time (15 ns typical) with respect to the edge of the logic trigger" referes to the amount of time in which the combination conditions of the logic input must be met before the detection of ...
item: 1225 | http://tmi.yokogawa.com/ar/service-support/faqs/1225/
How Long Does it Take for the DL2700 to Save Data to the External MO or Hard Disk via the SCSI?
The performance varies depending on the firmware version. The details are shown as follows: Saving byte data to a 230 MB MO: 1 kW:4 s 10 kW:4 s 50 kW:6 s 100 kW:7 s 250 kW:11 s 500 kW:18 s 1 MW:32 s 2 MW:59 ...
item: 1216 | http://tmi.yokogawa.com/ar/service-support/faqs/1216/
Why Won't the Files, Saved After Connecting a Hard Disk to the DL2700 via SCSI Interface, Appear on the PC?
The following procedure will allow the hard disk to be read properly. For Windows 95, open the "Settings" tab in the properties dialog box for the connected drive, and check the Removable box. Select "My Computer -> Con...
item: 1228 | http://tmi.yokogawa.com/ar/service-support/faqs/1228/
What is the Delay Time of the External Trigger Input on the DL2700?
There is zero delay because the external trigger circuit on the DL2700 is located before the A/D converter. The timing is approximately 10 ns faster than the trigger captured on the measured waveform, since there are no ...
item: 1226 | http://tmi.yokogawa.com/ar/service-support/faqs/1226/
Why Doesn't the DL2700 Stop Acquisition After One Acquisition When Single Start and Average Mode is Enabled?
During the averaging mode, the operation of the Single Start key differs from the usual operation. Exponential averaging: Operates in the same fashion as when the START/STOP key is pressed. The operation does not stop...
item: 1222 | http://tmi.yokogawa.com/ar/service-support/faqs/1222/