What driver should I use when connecting WT500/WT1800 via USB from NI LabVIEW Environment?
To use USB interface on the WT500 and WT1800 from NI LabVIEW environment, you will need to use the USB driver from National Instruments. This USB driver is usually installed when you install NI-VISA and is called the US...
item: 1313 | http://tmi.yokogawa.com/ar/service-support/faqs/1313/
Why do I Need to use a Hub When Communicating With the Instrument via Ethernet?
The AC Power Input in all Yokogawa instruments is designed as a 3-pin connection (one of which is a GND pin). In some parts of the world, PCs are sold with AC power cables that are 2-pin. Often times this means the PC d...
item: 906 | http://tmi.yokogawa.com/ar/service-support/faqs/906/
Power Analyzer Accuracy and Basic Uncertainty Calculator
The Power Analyzer Accuracy and Basic Uncertainty Calculator can be used to determine the uncertainty in voltage, current, and active power (watts) measurement values for various frequency ranges and wiring systems. To u...
item: 417 | http://tmi.yokogawa.com/ar/service-support/faqs/417/
Why Doesn't WTViewer Communicate with my WT Series Power Meter via GPIB?
There are several items you will need to check and verify to solve this issue. Verify the GP-IB connectionSome instruments have a D/A output connector located next to the GP-IB connector. There have been some instance...
item: 158 | http://tmi.yokogawa.com/ar/service-support/faqs/158/
Why Can't I Save the WT Settings for the WT500, WT1800, or WT3000 When Using WTViewer?
The WTViewer Save Setting feature will only save the setting information of the WTViewer. The Save Setting feature will NOT save the setting information of the instrument. The saved file for the WT500, WT1800 or WT3000 ...
item: 905 | http://tmi.yokogawa.com/ar/service-support/faqs/905/
Why is the Peak Value and Crest Factor Unstable?
The peak value and crest factor may be unstable if they have not been captured accurately. If the peak value is not stable, neither will the crest factor be stable. The cause is the difficulty in capturing the narrow spi...
item: 1264 | http://tmi.yokogawa.com/ar/service-support/faqs/1264/
Is the DA Output Terminal Connected to the Case of the WT Series Power Analyzer?
The WT3000 D/A output terminal is electrically isolated from the case. For all other models, the D/A output terminal is connected to the case. ...
item: 650 | http://tmi.yokogawa.com/ar/service-support/faqs/650/
Why are There Binding Posts on the Direct Current Input Terminals?
This is to prevent an open current circuit. Among non-Japanese power meters, there are products that also use safety terminals for current terminals. Safety terminals can be said to be safe because the terminal is not ex...
item: 1270 | http://tmi.yokogawa.com/ar/service-support/faqs/1270/
Can I Modify / Upgrade a WT500 to Add the EX Option?
Yes, please contact your nearest Yokogawa representative for more details. ...
item: 1055 | http://tmi.yokogawa.com/ar/service-support/faqs/1055/
What is the Part Number for the Current Knobs and Pads (Studder) Used on the Direct Input Terminals of the WT500?
The part numbers for the black current knobs and pads (studder) is listed below: A9105ZG: Black Current Knob Set of 2 B9292GX: Pad (Studder) Set of 10...
item: 940 | http://tmi.yokogawa.com/ar/service-support/faqs/940/