2406E SERIES OF ANALOG INSULATION TESTER

The 2406E Analog Insulation Tester:
page links
Details
Specifications - Model 240631
Standard Test Conditions - Model 240631
| Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
| Position of Use | Horizontal (5º max. of angle of inclination) |
| External Magnetic Fields | None Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
| Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |
Specifications - Model 240641
Standard Test Conditions - Model 240641
| Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
| Position of Use | Horizontal (5º max. of angle of inclination) |
| External Magnetic Fields | None Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
| Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |
Specifications - Model 240632
Standard Test Condtions - Model 240632
| Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
| Position of Use | Horizontal (5º max. of angle of inclination) |
| External Magnetic Fields | None Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
| Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |
Specifications - Model 240642
Standard Test Conditions - Model 240642
| Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
| Position of Use | Horizontal (5º max. of angle of inclination) |
| External Magnetic Fields | None Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
| Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |
Specifications - Model 240633
Standard Test Conditions - Model 240633
| Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
| Position of Use | Horizontal (5º max. of angle of inclination) |
| External Magnetic Fields | None Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
| Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |
Specifications - Model 240343
Standard Test Conditions - Model 240643
| Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
| Position of Use | Horizontal (5º max. of angle of inclination) |
| External Magnetic Fields | None Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
| Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |
Specifications - Model 240634
Standard Test Condtions - Model 240634
| Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
| Position of Use | Horizontal (5º max. of angle of inclination) |
| External Magnetic Fields | None Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
| Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |
Specifications - 240644
Standard Test Conditions - Model 240644
| Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
| Position of Use | Horizontal (5º max. of angle of inclination) |
| External Magnetic Fields | None Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
| Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |
Specifications - Model 240635
Standard Test Conditions - Model 240635
| Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
| Position of Use | Horizontal (5º max. of angle of inclination) |
| External Magnetic Fields | None Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
| Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |
Specifications - Model 240645
Standard Test Conditions - Model 240645
| Ambient Temperature/Humidity Ranges | 23 ±5 /45-75% RH |
| Position of Use | Horizontal (5º max. of angle of inclination) |
| External Magnetic Fields | None Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.) |
| Tolerance Mentioned Conditions | Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max. |
Model Numbers
| Model | Description |
|---|---|
| 240632 | Rating 125V/20M~ ; Effective Measuring Range 0.01-20M~; AC Voltage Measurement 0 to 300 V ; Lower Limit 0.125~; Rated Current 1 mA |
| 240641 | Rating 125V/20M~ 25V/5M~ 50V/10M~; Effective Measuring Range 0.001-5M~ 0.005-10M~ 0.01-20M~; AC Voltage Measurement 0 to 300 V ; Lower Limit 0.025M~ 0.05M~ 0.125~; Rated Current 1 mA |
| 240631 | Rating 125V/20M~ 25V/5M~ 50V/10M~; Effective Measuring Range 0.001-5M~ 0.005-10M~ 0.01-20M~; AC Voltage Measurement 0 to 300 V ; Lower Limit 0.025M~ 0.05M~ 0.125~; Rated Current 1 mA |
| 240634 | Rating 1000V/2000M~ 250V/50M~ 500V/100M~; Effective Measuring Range 0.01-0M~ 0.05-100M~ 1-2000M~; AC Voltage Measurement 0 to 600 V ; Lower Limit 0.25M~ 0.5M~ 1~; Rated Current 1 mA |
| 240644 | Rating 1000V/2000M~ 250V/50M~ 500V/100M~; Effective Measuring Range 0.01-0M~ 0.05-100M~ 1-2000M~; AC Voltage Measurement 0 to 600 V ; Lower Limit 0.25M~ 0.5M~ 1~; Rated Current 1 mA |
| 240635 | Rating 1000V/2000M~ 250V/50M~ 500V/100M~; Effective Measuring Range 0.01-0M~ 0.05-100M~ 1-2000M~; AC Voltage Measurement 0 to 600 V ; Lower Limit 0.25M~ 0.5M~ 1~; Rated Current 1 mA |
| 240645 | Rating 1000V/2000M~ 250V/50M~ 500V/100M~; Effective Measuring Range 0.01-0M~ 0.05-100M~ 1-2000M~; AC Voltage Measurement 0 to 600 V ; Lower Limit 0.25M~ 0.5M~ 1~; Rated Current 1 mA |
| 240643 | Rating 125V/20M~ 250V/50M~ 500V/100M~; Effective Measuring Range 0.01-20M~ 0.01-50M~ 0.05-100M~; AC Voltage Measurement 0 to 600 V ; Lower Limit 0.125M~ 0.25M~ 0.5~; Rated Current 1 mA |
| 240633 | Rating 125V/20M~ 250V/50M~ 500V/100M~; Effective Measuring Range 0.01-20M~ 0.01-50M~ 0.05-100M~; AC Voltage Measurement 0 to 600 V ; Lower Limit 0.125M~ 0.25M~ 0.5~; Rated Current 1 mA |
| 240642 | Rating 250V/50M~ ; Effective Measuring Range 0.01-50M~; AC Voltage Measurement 0 to 300 V ; Lower Limit 0.25; Rated Current 1 mA |