What is the Maximum Input Power for Pulsed Light Measurements on the AQ6370B?
The maximum input power that the AQ6370B can measure for pulsed light measurements is +20dBm per resolution at full span, per channel. Keep in mind that this limit is applicable to an averaged power of pulsed light, not ...
item: 161 | http://tmi.yokogawa.com/au/service-support/faqs/161/
Which eSATA and USB External Hard Drive is Compatible With the ScopeCorder?
The following eSATA and USB external hard drives are compatible with the ScopeCorder: Buffalo HD-HS500SU2 500GB IOMEGA LPHD080-U 80GB ELECOM LaCie d2 quadra LCH-2D1TQ 1TB IO DATA RHD-UX1.0T 1TB Logitec LHD-ED1000SAU...
item: 3 | http://tmi.yokogawa.com/au/service-support/faqs/3/
Is There a Way to Recover a DL850 WDF File That Cannot be Opened in Xviewer?
If the DL850 WDF file was saved through the hard disk recording feature (HD Recording), then the file may be recovered by using the Finalize feature in Xviewer File Utility. Files that were saved through the normal save ...
item: 1284 | http://tmi.yokogawa.com/au/service-support/faqs/1284/
How do I use the DLM2000 Enhanced Parameter Measurement and Calculation Feature?
The DLM2000 Series Oscilloscope is capable of performing enhanced parameter measurements and calculations. This feature allows you to perform automated measurements of waveform parameters of two different areas. It will...
item: 1024 | http://tmi.yokogawa.com/au/service-support/faqs/1024/
Why Can't the SL1000 Acquisition Software Detect my USB Connection?
Even though the correct USB drivers have been installed, and the device displays correctly in the Windows Device Manager Window, the SL1000 Acquisition Software fails to detect the SL1000 main unit. Solutions: Verify th...
item: 675 | http://tmi.yokogawa.com/au/service-support/faqs/675/
Does the Header File .HDR Saved by a ScopeCorder Contain the Instrument Settings?
Unfortunately, the header file .HDR saved by the ScopeCorder does not contain the instrument settings. If the .WVF and .HDR file pair was saved by the DL750/DL750P/SL1400, you may reload the WVF file back into the instru...
item: 315 | http://tmi.yokogawa.com/au/service-support/faqs/315/
What is the Equation Used to Convert the Voltage Axis Units From dBV to V When the Power Spectrum is Displayed on the DL1540?
If Y(V) is the converted value for X(dBV), the equation is: Y = 10(X/20) from X=20*LOG(Y) ...
item: 1191 | http://tmi.yokogawa.com/au/service-support/faqs/1191/
Why Does the Trigger Activate Repeatedly When the Detection Level is set to "H Active" and External Trigger Input is "H" for the WE7251/WE7271 Digitizer Module?
The external trigger input (TRIG IN) as well as external input/output (EXT. I/O) trigger signal becomes the level signal. The WE7251/WE7271 digitizer module will detect the bus trigger by that level. As long as the exte...
item: 1174 | http://tmi.yokogawa.com/au/service-support/faqs/1174/
How Long is the WE7251/WE7271/WE7272/WE7275 Digitizer Module's Record Length?
For high-speed data recording in trigger mode/gate mode, the digitizer module is equipped with internal memory to store measurement data in order to record reliable data without receiving any effects of the load conditio...
item: 1148 | http://tmi.yokogawa.com/au/service-support/faqs/1148/
How Much Lag Time is There When Inputting or Outting a Signal from the EXT Terminal of the WE7000 Measurement Station?
The lag time from the external terminal to the bus trigger / time base is as follows:When inputting from TRIG IN:TRIG IN -> bus trigger signal : 40 ns (typical value)TRIG IN -> time base signal : 25 ns (typical value)Whe...
item: 1139 | http://tmi.yokogawa.com/au/service-support/faqs/1139/