What is the Display Update Rate for the PZ4000?

The actual display update rate is shown below for observation times from 2 ms to 100 ms. 2 ms : 0.8 s 4 ms : 0.9 s 10 ms : 1.2 s 20 ms : 1.8 s 40 ms : 1.8 s 100 ms : 1.8 s Measurement Conditions Modules mounted : three...

item: 1180  |  http://tmi.yokogawa.com/bx/service-support/faqs/1180/

How do I set a Gage Factor Greater Than 2.2 for the 701271 Strain Input Module?

The Gage Factor setting for the 701271 Strain Input Module can be directly inputted within the range of 1.90 to 2.20. However, it is possible to indirectly set the Gage Factor to a larger value by setting the constant A ...

item: 569  |  http://tmi.yokogawa.com/bx/service-support/faqs/569/

How do I Update the Firmware on the SL1000?

Please follow the procedures listed in the attached document for instructions on how to update the firmware on the SL1000....

item: 329  |  http://tmi.yokogawa.com/bx/service-support/faqs/329/

Will the Zero Level Correction be Executed While in Auto-Range for the WT210/230?

Yes, if the instrument is in auto-range, zero level correction will still be executed. ...

item: 678  |  http://tmi.yokogawa.com/bx/service-support/faqs/678/

Is the Operating Temperature Range of the Modules the Same as the DL850/DL750 Main Chassis Unit?

Yes the operating temperature range of the individual modules is the same as the operating temperature range of the DL850/DL750 main chassis unit. ...

item: 1294  |  http://tmi.yokogawa.com/bx/service-support/faqs/1294/

How do I Retrieve Data From the Instrument Only When it is Updated Using IEEE488.2 Commands?

You can use the "STATUS:ESSR?" command to access the extended event register and determine whether the data was updated. You can judge the data update status by referencing bit 0 (UPD) of this register. However to do so,...

item: 1249  |  http://tmi.yokogawa.com/bx/service-support/faqs/1249/

Why Won't the Files, Saved After Connecting a Hard Disk to the DL2700 via SCSI Interface, Appear on the PC?

The following procedure will allow the hard disk to be read properly. For Windows 95, open the "Settings" tab in the properties dialog box for the connected drive, and check the Removable box. Select "My Computer -> Con...

item: 1228  |  http://tmi.yokogawa.com/bx/service-support/faqs/1228/

Xviewer 1.70 Math Issue

Xviewer 1.70 Math Issue Normally, for the math expression 'Mn' (when n...

item: 1342  |  http://tmi.yokogawa.com/bx/service-support/faqs/1342/

Is There any Information Related to the Durability of Luminanced of the LCD Monitor for the DL1540C/CL?

The manufacturer of our LCD has stated the luminance of the LCD drops to half after 25000 hours. ...

item: 1199  |  http://tmi.yokogawa.com/bx/service-support/faqs/1199/

Can I Remotely Control the DL850/DL850V Simultaneously using Ethernet and GPIB?

No, if you send commands simultaneously from another communications interface, that has not been selected on the DL850/DL850V, the instrument will not execute the commands properly. ...

item: 966  |  http://tmi.yokogawa.com/bx/service-support/faqs/966/