Why Doesn't the Power, Power Factor, and Phase Angle of Each Input Element Match When Measuring Balanced a Three-Phase Load?

Solution:

 In the case of three-phase three-wire, or 3V3A wiring, the power, power factor, and phase angle of each input element do not match because it is the line to line voltage that is measured.
 
Please download and refer to the attach PDF for a comprehensive explanation.