When Measuring the Balanced Three-Phase Load and Displaying the Waveform, the Phase Angle of the Voltage Across Each Input Element Does Not Equal 120 Degrees

In the three-phase three-wire, or 3V3A wiring scheme, the phase angle of voltage between each input element is 60 degrees because it is the line to line voltage that is measured.

Please download and refer to the attached PDF for more information.

Related Products & Solutions

Precision Power Analyzer WT3000

With 0.02% accuracy and 1MHz bandwidth, the WT3000 delivers where the highest precision measurements are required. It is the industry standard for R&D work on inverters, motor drives, lighting systems and electronic ballasts, UPS systems, aircraft power, transformer testing, and other power conversion devices.