KNOWLEDGEBASE

Filter your results by Category or Keyword

 

 

 

(29) Results were found. Showing 1-10.

I'm Thinking of Connecting the 751550 Clamp Probe to the External Shunt Input Terminal of the WT1000/WT2000. What Kind of Terminal Does the External Shunt Input use for the WT1000/WT2000 Series Instrument?

The WT1000/WT2000 series instrument uses a BNC connector for the external shunt input terminal. The output terminal of the 751550 Clamp Probe is a banana plug. Please use the 366921 Banana-to-BNC Conversion Adapter to co...

item: 1133  |  http://tmi.yokogawa.com/cl/service-support/faqs/1133/

Why are the Thermocouple Codes Shown on the Type 2553 DC Voltage/Current Standard Different From the Service Manual?

The specifications for the Type 2553 DC Voltage/Current Standard was changed on June 1982. The service manual was not rewritten to reflect these changes. There are two types of 2553:          1979/4 -- 1982/5         255...

item: 904  |  http://tmi.yokogawa.com/cl/service-support/faqs/904/

Power Analyzer Accuracy and Basic Uncertainty Calculator

The Power Analyzer Accuracy and Basic Uncertainty Calculator can be used to determine the uncertainty in voltage, current, and active power (watts) measurement values for various frequency ranges and wiring systems. To u...

item: 417  |  http://tmi.yokogawa.com/cl/service-support/faqs/417/

Why is the Peak Value and Crest Factor Unstable?

The peak value and crest factor may be unstable if they have not been captured accurately. If the peak value is not stable, neither will the crest factor be stable. The cause is the difficulty in capturing the narrow spi...

item: 1264  |  http://tmi.yokogawa.com/cl/service-support/faqs/1264/

Is the DA Output Terminal Connected to the Case of the WT Series Power Analyzer?

The WT3000 D/A output terminal is electrically isolated from the case. For all other models, the D/A output terminal is connected to the case. ...

item: 650  |  http://tmi.yokogawa.com/cl/service-support/faqs/650/

Why are There Binding Posts on the Direct Current Input Terminals?

This is to prevent an open current circuit. Among non-Japanese power meters, there are products that also use safety terminals for current terminals. Safety terminals can be said to be safe because the terminal is not ex...

item: 1270  |  http://tmi.yokogawa.com/cl/service-support/faqs/1270/

What is the Part Number for the Current Knobs and Pads (Studder) Used on the Direct Input Terminals of the WT500?

The part numbers for the black current knobs and pads (studder) is listed below: A9105ZG: Black Current Knob Set of 2 B9292GX: Pad (Studder) Set of 10...

item: 940  |  http://tmi.yokogawa.com/cl/service-support/faqs/940/

How do I Retrieve Measured Data Using Communications in Normal Mode?

Send the appropriate command as shown below to your instrument, then read in all the data that is returned.     PZ4000: "NUMERIC:NORMAL:VALUE?"    WT1600: "NUMERIC:NORMAL:VALUE?"    WT100/200 series instruments:        C...

item: 1240  |  http://tmi.yokogawa.com/cl/service-support/faqs/1240/

What is the Pinout for Ethernet and RS232 Cables?

Please download the attached PDF file for a list of pinouts for various communication cables....

item: 1075  |  http://tmi.yokogawa.com/cl/service-support/faqs/1075/

When Displaying the Harmonic Content of Each Order on the WT Series Power Analyzer, What Value is Used for the Denominator?

The value depends on the model of the power analyzer. For WT1000, WT2000, WT100, and WT200, it is fixed to the fundamental wave. For power analyzers with 7 segments LED, the relative harmonic content is fixed to the fu...