What is the Data Format of the ASCII File for the DL2700?
The data is saved in blocks of 8-bits and is expressed in decimal notation. In other words, if all 8-bits are zeroes, the decimal value is 0; if all 8-bits are 1s, the decimal value is 255....
item: 1218 | http://tmi.yokogawa.com/cl/service-support/faqs/1218/
Xviewer .FLD Floating Data Format Information
Xviewer .FLD Floating Data Format Information Xviewer saves wveform data as a .FLD Data ... here is example of how to gain access via VBA to that waveform data: Example1) Please copy the file "OK.FLD" to "C:temp"2) ...
item: 850 | http://tmi.yokogawa.com/cl/service-support/faqs/850/
What is the "Beat Method" for Inspecting an Oscilloscope’s Time Base Accuracy?
In the oscilloscope's QIS (Quality Inspection Standards), the "beat method" can be used to inspect and verify the time base accuracy. The "beat method" utilizes the aliasing effect to create a "beat waveform." The "beat ...
item: 449 | http://tmi.yokogawa.com/cl/service-support/faqs/449/
How do I Update the Voltage and Current Range on the WT100/WT200 Series Instrument Using Pre-IEEE488.2 Commands?
To change the voltage range on element 1 to 30 V, send the "CONFIGURE:VOLTAGE:RANGE 30V" command to the measuring instrument.To change the current range on element 1 to the 1 A range, send the "CONFIGURE:CURRENT:RANGE 1A...
item: 1252 | http://tmi.yokogawa.com/cl/service-support/faqs/1252/
Why Can't I Save the WT Settings for the WT500, WT1800, or WT3000 When Using WTViewer?
The WTViewer Save Setting feature will only save the setting information of the WTViewer. The Save Setting feature will NOT save the setting information of the instrument. The saved file for the WT500, WT1800 or WT3000 ...
item: 905 | http://tmi.yokogawa.com/cl/service-support/faqs/905/
Is VXI-11 Required for a DLM2000 to Xviewer Connection Over Ethernet?
Yes, VXI-11 is required and needs to be enabled for the DLM2000 to connect properly to Xviewer when using an ethernet connection. If the VXI-11 option is unchecked, a connection failed message will appear. Related Inf...
item: 208 | http://tmi.yokogawa.com/cl/service-support/faqs/208/
What is the Difference Between Type1 and Type2 With Regards to Reactive Power, Apparent Power, and Other Formulas?
Selecting formulas for calculating apparent power and reactive powerThere are several types of power—active power, reactive power, and apparent power. Generally, the following equations are satisfied:Active power P = UIc...
item: 1231 | http://tmi.yokogawa.com/cl/service-support/faqs/1231/
What is the maximum size of ATA Flash Card we can use with the WT3000?
What is the maximum size of ATA Flash Card we can use with the WT3000? The maximum size of the card to save data is 16GByte. The maximum size of the card used for firmware updating procss is limited to 256MByte.(A la...
item: 1407 | http://tmi.yokogawa.com/cl/service-support/faqs/1407/
DL850 ASCII Waveform Data File Hysteresis Offset
Using the DL850, the waveform data was directly saved as an ASCII file in the *.CSV and *.WDF file format (NOT an Xviewer conversion). The sample files that have been provided contain the same waveform data, which is a...
item: 1107 | http://tmi.yokogawa.com/cl/service-support/faqs/1107/
DL850 Time/Div Data Querying API Using Excel via USB Connection
The attached file contains an Excel spreadsheet with an API that will query data from the DL850 via USB, to obtain and display the Time/Div setting. To use the Excel API, follow these procedures: Download and extract t...
item: 990 | http://tmi.yokogawa.com/cl/service-support/faqs/990/