Can I Specify the Display Channel Color on the DL1540C/CL?

The display channel colors are fixed to:                 CH1-yellow, CH2-green, CH3-pink, CH4-blue....

item: 1197  |  http://tmi.yokogawa.com/de/service-support/faqs/1197/

Will all DL750 Modules Work on the DL850 or SL1000?

In most cases, all the DL750 modules will be compatible with the DL850 and SL1000. However, there are some module which can be used with DL750 but cannot be used with DL850. These are typically the early DL750 modules t...

item: 841  |  http://tmi.yokogawa.com/de/service-support/faqs/841/

Can the 702911 and 702912 be Used with the DL750 DL750P?

The 702911 and 702912 Logic Probes cannot be used with the DL750/DL750P. The logic probes were orginally designed to be used with the OR Series.   The SL1400 can use the 700986, 700987, 702911, 702912. The DL750 can onl...

item: 755  |  http://tmi.yokogawa.com/de/service-support/faqs/755/

Is There a Product Tutorial Guide for the SL1400?

Please see the attachment "Self-Paced Exercises for SL1400.PDF" for a set of easy exercises suitable for the SL1400 beginner. ...

item: 109  |  http://tmi.yokogawa.com/de/service-support/faqs/109/

Does the /JSxx Option Work for Xviewer Using Files Created on a DL750?

Unfortunately, it does not work for the files created on a DL750. It is a dedicated feature available only for the DL850/DL850V. ...

item: 802  |  http://tmi.yokogawa.com/de/service-support/faqs/802/

When Displaying the Harmonic Content of Each Order on the WT Series Power Analyzer, What Value is Used for the Denominator?

The value depends on the model of the power analyzer. For WT1000, WT2000, WT100, and WT200, it is fixed to the fundamental wave. For power analyzers with 7 segments LED, the relative harmonic content is fixed to the fu...

item: 642  |  http://tmi.yokogawa.com/de/service-support/faqs/642/

How do I set a Gage Factor Greater Than 2.2 for the 701271 Strain Input Module?

The Gage Factor setting for the 701271 Strain Input Module can be directly inputted within the range of 1.90 to 2.20. However, it is possible to indirectly set the Gage Factor to a larger value by setting the constant A ...

item: 569  |  http://tmi.yokogawa.com/de/service-support/faqs/569/

What is the Pinout for Ethernet and RS232 Cables?

Please download the attached PDF file for a list of pinouts for various communication cables....

item: 1075  |  http://tmi.yokogawa.com/de/service-support/faqs/1075/

How do I Update the Firmware on the SL1000?

Please follow the procedures listed in the attached document for instructions on how to update the firmware on the SL1000....

item: 329  |  http://tmi.yokogawa.com/de/service-support/faqs/329/

Why Can't the Width level of the Window Trigger on the DL1500 Series be set to Below the Voltage Sensitivity?

The width level of the Window trigger cannot be set to a value less than the voltage corresponding to 1 division at the specified voltage sensitivity. For example, if the voltage sensitivity is set to 500 V/div, then the...

item: 1200  |  http://tmi.yokogawa.com/de/service-support/faqs/1200/