KNOWLEDGEBASE

Item #: 1174   |   Created: 4/1/2013
Categories: Data Acquisition - WE7000

Return to Knowledgebase up pointing arrow

Why Does the Trigger Activate Repeatedly When the Detection Level is set to "H Active" and External Trigger Input is "H" for the WE7251/WE7271 Digitizer Module?

Solution:

The external trigger input (TRIG IN) as well as external input/output (EXT. I/O) trigger signal becomes the level signal. The WE7251/WE7271 digitizer module will detect the bus trigger by that level. As long as the external signals remain "true" or "high," the trigger will activate repeatedly.

Rate This Article

Document Quality?
Answered Your Question?

Webinar
How does the new WT300 provide reliable power measurements for your application?
View the webinar recording here.


Contact:
Yokogawa Europe BV

Tel.: +31 (0) 88 464 1429
Fax: +31 (0) 88 464 1107
tmi@nl.yokogawa.com