My AQ6317/AQ6317B OSA Locks up at Step 8 of 8 and will not Start Properly, How do I Fix This Issue?

When the AQ6317 /AQ6317B OSA stops at step 8 of 8 it usually means that the BIOS battery is dead. Please contact your nearest Yokogawa representative to have your unit sent in for repair and calibration. ...

item: 1060  |  http://tmi.yokogawa.com/es/service-support/faqs/1060/

When the Standard Operating Condition States "After 30 Minutes Warm-up and After Calibration," Does the "Calibration" Refer to the ScopeCorder's Auto-Cal Routine?

No, this "calibration" does not refer to the ScopeCorders Auto-Cal feature. When the Auto-Cal function is set to OFF, the ScopeCorder will not automatically calibrate after a 30 minute warm-up period. You must do a manua...

item: 1295  |  http://tmi.yokogawa.com/es/service-support/faqs/1295/

Xviewer 1.64 Will Not Open *.WDF in Realtime

Please make sure you have the latest version of Xviewer installed....

item: 1017  |  http://tmi.yokogawa.com/es/service-support/faqs/1017/

Is There a List of Compatible IRIG Generator for the DL850?

We have tested the following IRIG generator for compatibility with the DL850. GPS200 (Masterclock)http://www.masterclock.com/products/gps200.phphttp://www.masterclock.com/popup/standard_antenna.htm ES101 (ESE)http://ww...

item: 799  |  http://tmi.yokogawa.com/es/service-support/faqs/799/

Why WT210/230 and WT3000 can not calculate Urms, Unm, and Udc simultaneously?

There are restrictions of the DSP hardware.   WT230 and WT210 are low-cost power meters. So, a low-cost, modest performance DSP utilized.This DSP has slow calculation.   WT3000 is performing complicated calculation ...

item: 1377  |  http://tmi.yokogawa.com/es/service-support/faqs/1377/

What is the Part Number for the Current Knobs and Pads (Studder) Used on the Direct Input Terminals of the WT500?

The part numbers for the black current knobs and pads (studder) is listed below: A9105ZG: Black Current Knob Set of 2 B9292GX: Pad (Studder) Set of 10...

item: 940  |  http://tmi.yokogawa.com/es/service-support/faqs/940/

Is the Operating Temperature Range of the Modules the Same as the DL850/DL750 Main Chassis Unit?

Yes the operating temperature range of the individual modules is the same as the operating temperature range of the DL850/DL750 main chassis unit. ...

item: 1294  |  http://tmi.yokogawa.com/es/service-support/faqs/1294/

What is Meant by the "Setup Time (15 ns Typical) With Respect to the Edge of the Logic Trigger?"

The following statement "setup time (15 ns typical) with respect to the edge of the logic trigger" referes to the amount of time in which the combination conditions of the logic input must be met before the detection of ...

item: 1225  |  http://tmi.yokogawa.com/es/service-support/faqs/1225/

Why Can't the Width level of the Window Trigger on the DL1500 Series be set to Below the Voltage Sensitivity?

The width level of the Window trigger cannot be set to a value less than the voltage corresponding to 1 division at the specified voltage sensitivity. For example, if the voltage sensitivity is set to 500 V/div, then the...

item: 1200  |  http://tmi.yokogawa.com/es/service-support/faqs/1200/

Can Data be Saved in BMP Format to the Floppy Disk for DL1540 Models Without the Built-in Printer?

Yes, screen image data can be saved in BMP and other formats even if the built-in printer is not installed. ...

item: 1184  |  http://tmi.yokogawa.com/es/service-support/faqs/1184/