What are the Minimum Requirements for Linking Multiple SL1000 Units in Synchronization Mode?

The SL1000 is capable of linking up to 8 units, up to 128ch max, in synchronized mode using our SL1000 Acquisition Software. Using a common LINK file, data from all units can be processed and analyzed, as one, at the sa...

item: 306  |  http://tmi.yokogawa.com/es/service-support/faqs/306/

How do I Retrieve Data From the Instrument Only When it is Updated Using IEEE488.2 Commands?

You can use the "STATUS:ESSR?" command to access the extended event register and determine whether the data was updated. You can judge the data update status by referencing bit 0 (UPD) of this register. However to do so,...

item: 1249  |  http://tmi.yokogawa.com/es/service-support/faqs/1249/

Why Doesn't the DL2700 Stop Acquisition After One Acquisition When Single Start and Average Mode is Enabled?

During the averaging mode, the operation of the Single Start key differs from the usual operation. Exponential averaging: Operates in the same fashion as when the START/STOP key is pressed. The operation does not stop...

item: 1222  |  http://tmi.yokogawa.com/es/service-support/faqs/1222/

Why Does the Sampling Rate Display Disappear During FFT Analysis for the DL1500 Series?

The FFT is computed based on the ACQ data. In order to perform the FFT over 1 k points, the ACQ data are decimated (reduced) from 10 kW to 1 kW. At this point the sampling rate is equivalent to 100 kSs (1/10 of 1 MS/s). ...

item: 1196  |  http://tmi.yokogawa.com/es/service-support/faqs/1196/

Why Doesn't the Measured Value on the Main Unit Match the DA Output Displayed Value?

The following may be causing the problem. 5V may have occurred during rating. Check the range setting again. DA output error can affect the values when the input is smaller than the rating. Have you checked the error of...

item: 1267  |  http://tmi.yokogawa.com/es/service-support/faqs/1267/

What is the Difference Between the WE7271 and the WE7272 Digitizer Modules?

The input connectors are different, the WE7271 has a clamp type terminal, and the WE7272 has a BNC terminal. Also, the maximum common mode voltage, and maximum voltage between channels vary in their specifications betwe...

item: 1151  |  http://tmi.yokogawa.com/es/service-support/faqs/1151/

How Many Characters Can be Placed in a Waveform Label on the DL850?

You can input up to 16 characters in a waveform label, but the DL850 will only display 8 characters max. ...

item: 732  |  http://tmi.yokogawa.com/es/service-support/faqs/732/

Will Xviewer Display Waveform Data From Multiple DL850s and Synchronize the Data in Time?

Yes, Xviewer can be used to display the waveform data from multiple DL850's into a single window display with the data synchronized over time. ...

item: 720  |  http://tmi.yokogawa.com/es/service-support/faqs/720/

In Serial Bus Analysis Which Standard is it Necessary to Use a Differential Probe?

There are three kinds of standards for which a differential probe is necessary:  FlexRay CAN UART (Some but not all UART)  It is not necessary to use a differential probe for the LIN, I2C and SPI standards. ...

item: 407  |  http://tmi.yokogawa.com/es/service-support/faqs/407/

Why Won't the Numerical Value I Enter for Record Length Save for the WE7275/WE7271/WE7272 Digitizer Module?

The minimum record length which can be input is limited such that "record length x sampling interval" must be 5 milliseconds or more. When you set the record length to a value smaller than this it automatically changes t...

item: 1173  |  http://tmi.yokogawa.com/es/service-support/faqs/1173/


Webinar
How does the new WT300 provide reliable power measurements for your application?
View the webinar recording here.