What is Hi-Resolution Mode?
Solution:
High Resolution mode can be used to remove high frequency noise and increase vertical resolution. It achieves this by increasing the number of effective bits per data to 12-bits through digital and bandwidth filters. The algorithm it uses to achieve the 12-bits of vertical resolution is:
Pre-Filter ⇒Over-Sample ⇒Average ⇒Data-Thinning ⇒Filter
Please download and review the attached PDF for a comprehensive overview of vertical resolustion and high resolution mode for the DL series oscilloscopes.