Why Doesn't the WT3000 Display the Phase Difference φU1-U2 of Voltage Between Phases?
Please verify the following settings on the WT3000: Wiring System set to 3P3W or 3V3A ITEM set to φUi-Uj Element set to ΣA ITEM set to φUi-UkWiring 3P4W or 3P3W(3V3A)Assign element 1,2, 3 to ΣA Update rate is not too f...
item: 596 | http://tmi.yokogawa.com/fr/service-support/faqs/596/
Can I Make a Direct One-to-One Connection From the PC to the Measuring Station Using a Cross Cable?
Connection made by a cross cable does not comply with Ethernet standards. For reliable data communications, please use a hub when you want to make a direct connection to a PC. ...
item: 1158 | http://tmi.yokogawa.com/fr/service-support/faqs/1158/
Is There an Automated Adjustment Procedure Available if the DL9000 is Failing Calibration?
Unfortunately we do not have an automated program for adjusting the DL9000 is it is failing calibration. Please follow the procedures on the DL9000 service manual or the CE handbook. ...
item: 423 | http://tmi.yokogawa.com/fr/service-support/faqs/423/
Can I use an AQ9335 Adapter Instead of an AQ9335C Adapter on the AQ2200 Sensor Modules?
No, although the basic shape is the same for both the AQ9335 and AQ9335C, the internal physical deisgn for the AQ9335C has been changed slightly to fit the AQ2200 series sensor modules for optimum accuracy. You will need...
item: 225 | http://tmi.yokogawa.com/fr/service-support/faqs/225/
What is the Warranty Period of Differential Probes ?
The warranty period of the differential probes, 701920, 701922, 701923 and 701924, are 1 year. If the probe is delivered from Yokogawa Japan, the warranty certificate of Yokogawa would be in the box that the probe was de...
item: 101 | http://tmi.yokogawa.com/fr/service-support/faqs/101/
Why are the Measured Values Negative When Measuring a Balanced Three-Phase Load?
In the three-phase three-wire, or 3V3A wiring scheme, the phase angle of voltage and current input to each input differs from that of the actual load because it is the line to line voltage that is measured. In three-pha...
item: 1261 | http://tmi.yokogawa.com/fr/service-support/faqs/1261/
Is it Possible to AC Couple the Input With the Digitizer Module?
Yes, AC coupling is supported with the WE7275 2-CH, 1 MS/s Isolated digitizer module. ...
item: 1147 | http://tmi.yokogawa.com/fr/service-support/faqs/1147/
Is There a List of Compatible USB Printers for the AQ7270 and AQ7275 Series OTDR?
Please download the attached Word document for a list of compatible USB drivers for the AQ7270 and AQ7275 series OTDR....
item: 1111 | http://tmi.yokogawa.com/fr/service-support/faqs/1111/
How do I Update the Voltage and Current Range on the Instrument Using Pre-IEEE488.2 Command Types?
To change the voltage range on element 1 to the 30 V range, send the "RV1,4" command to the measuring instrument. To change the current range on element 1 to the 1 A range, send the "RA1,5" command to the measuring inst...
item: 1242 | http://tmi.yokogawa.com/fr/service-support/faqs/1242/
How do I Achieve 1usec Resolution Timestamps on the DLM2000 History List?
The DLM2000 has a unique feature that will allow you to save the timestamps in the History List. Please verify or update the firmware version on your DLM2000 to version 1.81 or later. There are two different save options...
item: 1106 | http://tmi.yokogawa.com/fr/service-support/faqs/1106/