720220 16-Channel Voltage Module Repeating Data Behavior for the DL850
If the only active module in the DL850 Chassis is the 720220 16 Channel Voltage Module, the waveform data file (ASCII or WDF) will NOT contain repeated samples. If multiple modules, including the 720220, are installed in...
item: 1088 | http://tmi.yokogawa.com/gb/service-support/faqs/1088/
What is Hi-Resolution Mode?
High Resolution mode can be used to remove high frequency noise and increase vertical resolution. It achieves this by increasing the number of effective bits per data to 12-bits through digital and bandwidth filters. Th...
item: 1076 | http://tmi.yokogawa.com/gb/service-support/faqs/1076/
Why Can't I Output the D/A Output of the Voltage and Current Frequency Correctly?
The frequency display and D/A output can handle only one input element at a time, either the voltage or the current. When either the V Hz or A Hz display function indicators are lit in display C, the frequency correspond...
item: 1237 | http://tmi.yokogawa.com/gb/service-support/faqs/1237/
What is the Data Format of the ASCII File for the DL2700?
The data is saved in blocks of 8-bits and is expressed in decimal notation. In other words, if all 8-bits are zeroes, the decimal value is 0; if all 8-bits are 1s, the decimal value is 255....
item: 1218 | http://tmi.yokogawa.com/gb/service-support/faqs/1218/
I'm Thinking of Connecting the 751550 Clamp Probe to the External Shunt Input Terminal of the WT1000/WT2000. What Kind of Terminal Does the External Shunt Input use for the WT1000/WT2000 Series Instrument?
The WT1000/WT2000 series instrument uses a BNC connector for the external shunt input terminal. The output terminal of the 751550 Clamp Probe is a banana plug. Please use the 366921 Banana-to-BNC Conversion Adapter to co...
item: 1133 | http://tmi.yokogawa.com/gb/service-support/faqs/1133/
When Performing a "ZONE" Search Using DL750 History, if "No Matching Record" is Found in History, the DL750 Displays a Warning Message. Is There a Programmatic Method to Check/Read This Warning Message?
Unfortunately there is no way to read these warning messages on the DL750. In addition, there is no other command available to determine if there was "No Matching Record" found in History for a particular ZONE condition...
item: 1082 | http://tmi.yokogawa.com/gb/service-support/faqs/1082/
How do I Achieve Higher Resolution Timestamps When Viewing ASCII Files?
If you wish to view the timestamp data in high resolution, you will need to open the *.CSV file with Notepad or Wordpad. If you use Microsoft Excel to open the file, Excel will round the timestamp data, preventing you ...
item: 1110 | http://tmi.yokogawa.com/gb/service-support/faqs/1110/
Why do I Need to use a Hub When Communicating With the Instrument via Ethernet?
The AC Power Input in all Yokogawa instruments is designed as a 3-pin connection (one of which is a GND pin). In some parts of the world, PCs are sold with AC power cables that are 2-pin. Often times this means the PC d...
item: 906 | http://tmi.yokogawa.com/gb/service-support/faqs/906/
Is the Data Assignment the Same When the Logic Input Values are Saved in ASCII Format on the DL708/DL708E and the DL2700?
DL708/DL708E: The ON/OFF conditions of bit signals from A1 to A8 and B1 to B8 are expressed as a 16-bit value. The 16-bit value is then converted to a decimal value and the result is saved. The value is written as a numb...
item: 1220 | http://tmi.yokogawa.com/gb/service-support/faqs/1220/
Can the External Clock and External Trigger Functions on the DL1500 Series be Specified or Used Simultaneously?
Yes, there is no restriction on specifying or using the two functions simultaneously. ...
item: 1201 | http://tmi.yokogawa.com/gb/service-support/faqs/1201/