(27) Results were found. Showing 1-10.
Power Analyzer Accuracy and Basic Uncertainty Calculator
The Power Analyzer Accuracy and Basic Uncertainty Calculator can be used to determine the uncertainty in voltage, current, and active power (watts) measurement values for various frequency ranges and wiring systems. To u...
item: 417 | http://tmi.yokogawa.com/gb/service-support/faqs/417/
Why Doesn't WTViewer Communicate with my WT Series Power Meter via GPIB?
There are several items you will need to check and verify to solve this issue. Verify the GP-IB connectionSome instruments have a D/A output connector located next to the GP-IB connector. There have been some instance...
item: 158 | http://tmi.yokogawa.com/gb/service-support/faqs/158/
Why is the Theoretical Calculated Power Factor and Phase Angle Different Than the Measured Values?
When measuring input signals of distorted waves, signals that are DC-offset or signals that include superimposed harmonic components, will result in different values for power factor and phase angle than those expected w...
item: 1263 | http://tmi.yokogawa.com/gb/service-support/faqs/1263/
How do I Update the Voltage and Current Range for the PZ4000 and WT1600 Using Communication Commands?
To change the voltage range on element 1 to 30 V, send the "INPUT:VOLTAGE:RANGE:ELEMENT1 30V" command to the measuring instrument.To change the current range on element 1 to the 1 A range, send the "INPUT:CURRENT:RANGE:...
item: 1250 | http://tmi.yokogawa.com/gb/service-support/faqs/1250/
How do I Retrieve Specific Items From the PZ4000 and WT1600 Using the Communications Commands?
Send the "NUMERIC:FORMAT:ASCII" command This sets the data format for the data you want to read out. Measured data read out using the "NUMERIC:NORMAL:VALUE?" command is output as an ASCII string. Send the "NUMERIC:NORM...
item: 1244 | http://tmi.yokogawa.com/gb/service-support/faqs/1244/
Why are There Binding Posts on the Direct Current Input Terminals?
This is to prevent an open current circuit. Among non-Japanese power meters, there are products that also use safety terminals for current terminals. Safety terminals can be said to be safe because the terminal is not ex...
item: 1270 | http://tmi.yokogawa.com/gb/service-support/faqs/1270/
How do I Retrieve Data From the PZ400 Only When it is Updated on the Instrument?
You can use the "STATUS:ESSR?" command to access the extended event register and determine whether the data was updated. You can judge the data update status by referencing bit 1 (DAV) of this register. However to do so,...
item: 1239 | http://tmi.yokogawa.com/gb/service-support/faqs/1239/
How do I Analyze the Waveform Data Saved From the PZ4000?
You can use the Power Viewer (Model 253734) software. However if you only need to view waveforms, we recommend you use the Waveform Viewer (Model 700919, version 1.23 or later) software. The trial version of the Power V...
item: 1179 | http://tmi.yokogawa.com/gb/service-support/faqs/1179/
What is the Part Number for the Current Knobs and Pads (Studder) Used on the Direct Input Terminals of the WT500?
The part numbers for the black current knobs and pads (studder) is listed below: A9105ZG: Black Current Knob Set of 2 B9292GX: Pad (Studder) Set of 10...
item: 940 | http://tmi.yokogawa.com/gb/service-support/faqs/940/
When Measuring the Balanced Three-Phase Load and Displaying the Waveform, the Phase Angle of the Voltage Across Each Input Element Does Not Equal 120 Degrees
In the three-phase three-wire, or 3V3A wiring scheme, the phase angle of voltage between each input element is 60 degrees because it is the line to line voltage that is measured. Please download and refer to the attache...
item: 1260 | http://tmi.yokogawa.com/gb/service-support/faqs/1260/