Why Doesn't the DL2700 Stop Acquisition After One Acquisition When Single Start and Average Mode is Enabled?

Solution:

During the averaging mode, the operation of the Single Start key differs from the usual operation.

  • Exponential averaging: Operates in the same fashion as when the START/STOP key is pressed. The operation does not stop until the START/STOP key is pressed.
  • Linear averaging: Stops after making measurements for the specified averaging count.