Solution:

Unbalanced, non-isolated inputs are single-ended amplifiers. These inputs are suitable for measurements where common mode voltages are extremely small or zero and measurements are referenced to GND.
 
Unbalanced, isolated inputs are single-ended, floating common amplifiers. These inputs are suitable for off-ground measurements up to the breakdown voltage of the isolation barrier. The common noise rejection ratio is modest at low frequencies.
 
Balanced, differential inputs are non-isolated and exhibit the best common noise rejection ratio at low and high frequencies. 

Please download and refer to the attached PowerPoint for more information on input types.

Rate This Article

Document Quality?
Answered Your Question?