To change the voltage range on element 1 to the 30 V range, send the "RV1,4" command to the measuring instrument.
To change the current range on element 1 to the 1 A range, send the "RA1,5" command to the measuring instrument.
To set the ranges on all elements at once, use: Voltage global setting "RV0,4" or Current global setting "RA0,5"
After entering settings, confirm the settings by sending the following commands:
For more information on the correspondence between voltage values, current values, and parameters, please refer to:
The WT2000 digital powermeter series has been designed with emphasis on basic performance (bandwidth, accuracy, response speed, and noise immunity) from the viewpoint of measurement of electrical quantities. The broad range of functions of these power analyzers enable them to be used in various fields of applications.