INTEG1 vs INTEG2 - as demonstrated on DL850 G3

INTEG1 vs INTEG2 - as demonstrated on DL850 G3   Examples of two different ways to INTEGRATE on the DL850/G3 DSP.   One attachment. ...

item: 1311  |  http://tmi.yokogawa.com/it/service-support/faqs/1311/

Is it Possible to Bypass the Router (Subnet) and Connect to the Measuring Station with the Ethernet Module?

Yes, you can bypass the router (subnet) and connect using the WE7000 control software ver. 4 and later. However, this is limited to the WE7052 Fast Ethernet module....

item: 1157  |  http://tmi.yokogawa.com/it/service-support/faqs/1157/

Can I use the Store Function to Record Waveform Data on the WT1800?

THE WT1800 and WT500 STORE function can only be used to record NUMERIC data.   The WT1600 and WT3000 STORE function can be used to record both NUMERIC and WAVEFORM data. ...

item: 915  |  http://tmi.yokogawa.com/it/service-support/faqs/915/

Why Can't I set the DL850/DL850V to 100 ns/div?

The DL850/DL850V cannot be set to 100ns/div when the following modules are installed: 720240 CAN Bus Monitoring Module 720220 16 Channel Voltage Input Module Removing these modules permits the DL850/DL850V to be adjus...

item: 845  |  http://tmi.yokogawa.com/it/service-support/faqs/845/

What is the Equation Used for the FFT Analysis on the DL1540?

It is the power spectrum defined by PS = 20 x Log(X/1Vrms) ...

item: 1190  |  http://tmi.yokogawa.com/it/service-support/faqs/1190/

How Can I Programatically Verify if the Calculations for Waveform Parameters are Completed?

There are two methods to programmatically determine if the calculations for waveform parameters is completed on the DL series oscilloscopes. Send the :MEAS:WAIT? command. This will ensure that no other commands are exec...

item: 1041  |  http://tmi.yokogawa.com/it/service-support/faqs/1041/

How can I Improve the Signal Quality of the G3 Frequency Function on the DL850?

For the DL850 G3 Frequency Function, you will need to set the hysterisis of the edges that are used to detect the periods. This will increase the tolerance for noise and improve measurement reliability.To configure the ...

item: 1039  |  http://tmi.yokogawa.com/it/service-support/faqs/1039/

Are LC connectors available for the AQ7275?

The SU2005A-LCC universal connector is physically compatible with AQ7275, but we do not recommend it because we expect poor connection repeatability and return loss. We cannot guarantee the performance of AQ7275 when the...

item: 667  |  http://tmi.yokogawa.com/it/service-support/faqs/667/

What is the Maximum Input Power for Pulsed Light Measurements on the AQ6370B?

The maximum input power that the AQ6370B can measure for pulsed light measurements is +20dBm per resolution at full span, per channel. Keep in mind that this limit is applicable to an averaged power of pulsed light, not ...

item: 161  |  http://tmi.yokogawa.com/it/service-support/faqs/161/

Which eSATA and USB External Hard Drive is Compatible With the ScopeCorder?

The following eSATA and USB external hard drives are compatible with the ScopeCorder: Buffalo HD-HS500SU2 500GB IOMEGA LPHD080-U 80GB ELECOM LaCie d2 quadra LCH-2D1TQ 1TB IO DATA RHD-UX1.0T 1TB Logitec LHD-ED1000SAU...

item: 3  |  http://tmi.yokogawa.com/it/service-support/faqs/3/