KNOWLEDGEBASE

Item #: 1251   |   Created: 4/1/2013
Categories: Power Meter - WT2030/1030/2533

Return to Knowledgebase up pointing arrow

How do I Update the Voltage and Current Range on the WT1000/WT2000 Series Instrument Using IEEE488.2 Commands?

Solution:

To change the voltage range on element 1 to 30 V, send the "CONFIGURE:VOLTAGE:RANGE:ELEMENT1 30V" command to the measuring instrument.
To change the current range on element 1 to the 1 A range, send the "CONFIGURE:CURRENT:RANGE:ELEMENT1 1A" command to the measuring instrument.

After entering settings, confirm the settings by sending the following commands:

  • To query the voltage on element 1: "CONFIGURE:VOLTAGE:RANGE:ELEMENT1?"
  • To query the current on element 1: "CONFIGURE:CURRENT:RANGE:ELEMENT1?" 

To set/query the ranges on all elements at once, use the following commands:

  1. Voltage global setting
    "CONFIGURE:VOLTAGE:RANGE 30V"
  2. Voltage global query
    "CONFIGURE:VOLTAGE:RANGE?"
  3. Current global setting
    "CONFIGURE:CURRENT:RANGE 1A"
  4. Current global query
    "CONFIGURE:CURRENT:RANGE?"

For information on the voltage and current values that can be set using these commands, please refer to:

  • WT1010 Digital Power Meter User's Manual
    Pages App2-20 and App2-22
  • WT1030/WT1030M Digital Power Meter User's Manual
    Pages App2-20 and App2-22
  • WT2010 Digital Power Meter User's Manual
    Pages App2-21 and App2-23
  • WT2030 Digital Power Meter User's Manual
    Pages App2-21 and App2-23