Why are There Binding Posts on the Direct Current Input Terminals?
This is to prevent an open current circuit. Among non-Japanese power meters, there are products that also use safety terminals for current terminals. Safety terminals can be said to be safe because the terminal is not ex...
item: 1270 | http://tmi.yokogawa.com/service-support/faqs/1270/
Which Bits in the Extended Event Register Corresponds to the “Check Range” LED Indicator on the WT210/WT230?
If you want to know the condition of the "Check Range" LED indicator for voltage and current, you will need to read three bits in the Extended Event Register. "Bit-6", "Bit-7", and "Bit-8" needs to be read to know the st...
item: 1095 | http://tmi.yokogawa.com/service-support/faqs/1095/
Why WT210/230 and WT3000 can not calculate Urms, Unm, and Udc simultaneously?
There are restrictions of the DSP hardware. WT230 and WT210 are low-cost power meters. So, a low-cost, modest performance DSP utilized.This DSP has slow calculation. WT3000 is performing complicated calculation ...
item: 1377 | http://tmi.yokogawa.com/service-support/faqs/1377/
Why Can't I Output the D/A Output of the Voltage and Current Frequency Correctly?
The frequency display and D/A output can handle only one input element at a time, either the voltage or the current. When either the V Hz or A Hz display function indicators are lit in display C, the frequency correspond...
item: 1237 | http://tmi.yokogawa.com/service-support/faqs/1237/
Is There a Way to Lock Out All the Keys, Including LOCAL, on the WT210/WT230?
It is possible to lock out all the keys, including LOCAL, on the WT210/WT230 using a GPIB specific or controller specific command.For the GPIB controller from National Instruments, the command is called LLO or Local Lock...
item: 1045 | http://tmi.yokogawa.com/service-support/faqs/1045/
What is the Part Number for the Current Knobs and Pads (Studder) Used on the Direct Input Terminals of the WT500?
The part numbers for the black current knobs and pads (studder) is listed below: A9105ZG: Black Current Knob Set of 2 B9292GX: Pad (Studder) Set of 10...
item: 940 | http://tmi.yokogawa.com/service-support/faqs/940/
Is WTViewer Supported Under the Linux Environment?
Although WTViewer is not officially supported under the Linux environment, users have successfully done so using WINE (flavor of Linux) via RS232. For connectivity to WT210/WT230, WTViewer requires that the meter be set ...
item: 1038 | http://tmi.yokogawa.com/service-support/faqs/1038/
How do I Retrieve Specific Items from the WT100/WT200 Using IEEE488.2 Commands?
Send the "MEASURE:NORMAL:ITEM:PRESET:CLEAR" command to the measuring instrument. This command turns all items for output OFF. Therefore no items will be output if you send the "MEASURE:NORMAL:VALUE?" command. Send the ...
item: 1247 | http://tmi.yokogawa.com/service-support/faqs/1247/
How do I Retrieve Specific Items From the WT100/WT200 Using Pre-IEEE488.2 Commands?
Send the "OF1,1,1" commandThis command sets the voltage value (V) from element 1 to output to channel 1 (from among the 14 output channels), and causes the voltage of element 1 to be read out. In the same manner, make...
item: 1246 | http://tmi.yokogawa.com/service-support/faqs/1246/
What are the Part Numbers for the 2 pcs Flat-Head and Binding-Head Screws in the WT210 Rack Mount Kit?
The part numbers for the 2 piece screws in the WT210 Rack Mount Kit (751533-E2) are listed below: Y9414LB: Binding-head screw (M4, 14mm long) Y9414EB: Flat-head screw (M4,14mm long) If you wish to purchase the screws fr...
item: 914 | http://tmi.yokogawa.com/service-support/faqs/914/