Are LC connectors available for the AQ7275?

The SU2005A-LCC universal connector is physically compatible with AQ7275, but we do not recommend it because we expect poor connection repeatability and return loss. We cannot guarantee the performance of AQ7275 when the...

item: 667  |  http://tmi.yokogawa.com/service-support/faqs/667/

Can the Width of the Harmonic Analysis Window be Increased Beyond 16?

In order to analyze the harmonics by the PLL synchronization, the maximum possible window width is automatically determined depending on the frequency of the fundamental wave of the input signal (fundamental frequency f)...

item: 1135  |  http://tmi.yokogawa.com/service-support/faqs/1135/

Is the Data Assignment the Same When the Logic Input Values are Saved in ASCII Format on the DL708/DL708E and the DL2700?

DL708/DL708E: The ON/OFF conditions of bit signals from A1 to A8 and B1 to B8 are expressed as a 16-bit value. The 16-bit value is then converted to a decimal value and the result is saved. The value is written as a numb...

item: 1220  |  http://tmi.yokogawa.com/service-support/faqs/1220/

Error:Code 915 Appears When Attempting to Print With the DL1540

Error:Code 915 is the error message "Can't detect listener." This error occurs when the hard copy destination is not set to the "printer," but to "plotter" without connecting external devices....

item: 1206  |  http://tmi.yokogawa.com/service-support/faqs/1206/

How Much Lag Time is There When Inputting or Outting a Signal from the EXT Terminal of the WE7000 Measurement Station?

The lag time from the external terminal to the bus trigger / time base is as follows:When inputting from TRIG IN:TRIG IN -> bus trigger signal : 40 ns (typical value)TRIG IN -> time base signal : 25 ns (typical value)Whe...

item: 1139  |  http://tmi.yokogawa.com/service-support/faqs/1139/

Why Does the Trigger Activate Repeatedly When the Detection Level is set to "H Active" and External Trigger Input is "H" for the WE7251/WE7271 Digitizer Module?

The external trigger input (TRIG IN) as well as external input/output (EXT. I/O) trigger signal becomes the level signal. The WE7251/WE7271 digitizer module will detect the bus trigger by that level. As long as the exte...

item: 1174  |  http://tmi.yokogawa.com/service-support/faqs/1174/

What does the Cursor Value on the AQ7275 Measure?

The cursor value displays the level (dB) of the signal that is above the noise floor of the connected optical port. The noise floor is the dynamic range. This lets you know if you are approaching the maximum distance tha...

item: 1054  |  http://tmi.yokogawa.com/service-support/faqs/1054/

What are the Minimum Configuration Requirements for the WT3000 to Work with the Flicker Software?

To use the WT3000 with the Flicker Software, the WT3000 must have the following options: /G6 option - Advanced Calculations /FL option - Flicker Measurements 1 to 3 elements - 30A Input Module Note: The 2A input mod...

item: 1051  |  http://tmi.yokogawa.com/service-support/faqs/1051/

Error(Code:846) Occurs When Attempting to Start Integration on the WT3000

"Error(Code:846): Attempted to start integration while measurement of peak overflow was in progress"The following error code will appear if Peak Over has occurred and you attempt to start integration. The WT3000 has two...

item: 805  |  http://tmi.yokogawa.com/service-support/faqs/805/

Is There an Arbitrary Waveform Data Creation Tool for Output on the WE7281 and WE7121 Module?

Yes, there is a software called the 707751 Arbitrary Waveform Editor. You can create waveforms using the editor's standard functions or with the mouse. You can read in files created in CSV, or in Yokogawa's proprietary ...

item: 1164  |  http://tmi.yokogawa.com/service-support/faqs/1164/