Can I Input Tens to Hundreds of Volts to the A/D Modules for the WE7000 Series?
The WE7241 digital temperature measurement module can measure up to ±50V. The WE7271/WE7272 4-CH, 100 kS/s Isolation digitizer module up to ±35V. The WE7275 2-CH, 1 MS/s Digitizer module accepts up to ±350V....
item: 1141 | http://tmi.yokogawa.com/kr/service-support/faqs/1141/
Can I Swap the WE7033/WE7034 Optical Interface Card for ISA bus to the WE7035/WE7036 PCI Optical Interface Card?
Yes, however the communication drivers are different so you will need a new driver for the WE7035/WE7036 optical interface card. The driver is located on the WE7000 Control Software CD ROM Ver. 3.1.2 and later or you can...
item: 1171 | http://tmi.yokogawa.com/kr/service-support/faqs/1171/
Why Can't I See the Harmonic Bar Graph in WTViewer After a Single Update When Connected to the WT1600?
For the WT1600, to view the harmonic data in the Bar Graph window of WTViewer, you have to select all the harmonic orders in the Numeric window. WTViewer's Bar Graph window display follows the harmonic items activated in...
item: 941 | http://tmi.yokogawa.com/kr/service-support/faqs/941/
Can a DL9000 MSO be Upgraded or Modified to a SB5000?
No, the DL9000 cannot be upgraded or modified to a SB5000....
item: 454 | http://tmi.yokogawa.com/kr/service-support/faqs/454/
What is the Output Resolution of the D/A Analog Option for the WT3000?
The output resolution for the WT3000 is 16-bits. ...
item: 1229 | http://tmi.yokogawa.com/kr/service-support/faqs/1229/
What is the Difference Between the DL850 and DL850V?
The only difference is that the CAN bus monitor module works with the DL850V only and the front panel sheet color is different....
item: 702 | http://tmi.yokogawa.com/kr/service-support/faqs/702/
Why Does the Wavelength Calibration Fail but Optical Alignment Pass on the AQ6375?
Please clean the fiber used between the calibration output and the optical input. ...
item: 507 | http://tmi.yokogawa.com/kr/service-support/faqs/507/
Why Won't my USB Flash Drive Read/Write on the DLM2000?
Please verify the following items: USB does not exceed 16 GB Latest firmware is installed on DLM2000 USB Format is FAT32 USB works properly on PC Test USB on another USB port If the problem persists, please contact Yok...
item: 1293 | http://tmi.yokogawa.com/kr/service-support/faqs/1293/
What is the Equation Used for the FFT Analysis on the DL1540?
It is the power spectrum defined by PS = 20 x Log(X/1Vrms) ...
item: 1190 | http://tmi.yokogawa.com/kr/service-support/faqs/1190/
Will Xviewer Display Waveform Data From Multiple DL850s and Synchronize the Data in Time?
Yes, Xviewer can be used to display the waveform data from multiple DL850's into a single window display with the data synchronized over time. ...
item: 720 | http://tmi.yokogawa.com/kr/service-support/faqs/720/