What is the Output Resolution of the D/A Analog Option for the WT3000?

The output resolution for the WT3000 is 16-bits. ...

item: 1229  |  http://tmi.yokogawa.com/kr/service-support/faqs/1229/

Why Does the Trigger Activate Repeatedly When the Detection Level is set to "H Active" and External Trigger Input is "H" for the WE7251/WE7271 Digitizer Module?

The external trigger input (TRIG IN) as well as external input/output (EXT. I/O) trigger signal becomes the level signal. The WE7251/WE7271 digitizer module will detect the bus trigger by that level. As long as the exte...

item: 1174  |  http://tmi.yokogawa.com/kr/service-support/faqs/1174/

Why Won't the Numerical Value I Enter for Record Length Save for the WE7275/WE7271/WE7272 Digitizer Module?

The minimum record length which can be input is limited such that "record length x sampling interval" must be 5 milliseconds or more. When you set the record length to a value smaller than this it automatically changes t...

item: 1173  |  http://tmi.yokogawa.com/kr/service-support/faqs/1173/

How Much Lag Time is There When Inputting or Outting a Signal from the EXT Terminal of the WE7000 Measurement Station?

The lag time from the external terminal to the bus trigger / time base is as follows:When inputting from TRIG IN:TRIG IN -> bus trigger signal : 40 ns (typical value)TRIG IN -> time base signal : 25 ns (typical value)Whe...

item: 1139  |  http://tmi.yokogawa.com/kr/service-support/faqs/1139/

How do I Monitor the Start/Stop Status of the DL716 Using Communication Commands?

It is possible to monitor the START/STOP status of the DL716 using communication commands. To do this, you will have to send the ":STATus:CONDition?" (without quotations) command to the instrument, in order to query the ...

item: 1050  |  http://tmi.yokogawa.com/kr/service-support/faqs/1050/

What is the Equation Used for the FFT Analysis on the DL1540?

It is the power spectrum defined by PS = 20 x Log(X/1Vrms) ...

item: 1190  |  http://tmi.yokogawa.com/kr/service-support/faqs/1190/

Is it Possible to Bypass the Router (Subnet) and Connect to the Measuring Station with the Ethernet Module?

Yes, you can bypass the router (subnet) and connect using the WE7000 control software ver. 4 and later. However, this is limited to the WE7052 Fast Ethernet module....

item: 1157  |  http://tmi.yokogawa.com/kr/service-support/faqs/1157/

How do I Programmatically Read Average Active Power Right Before the End of the Integration Interval While in Continuous Integration for the WT230?

You can read the average active power during continuous integration mode (just before the integral resets) for the WT230, by monitoring the status of the ITG or ITM bits of the extended event register.   Bit 1 ITG is u...

item: 982  |  http://tmi.yokogawa.com/kr/service-support/faqs/982/

What is the DL850 Realtime File Divide Feature?

The DL850 realtime file divide feature enables you to record long waveform acquisitions (realtime) divided into blocks of smaller files. Each file block can be 2GB in file size. You can specify the number of files that t...

item: 721  |  http://tmi.yokogawa.com/kr/service-support/faqs/721/

What is the On Start Trigger Mode on the DL850?

The On Start Trigger Mode (Log Mode for the DL750) will perform a single waveform acqusition immediately when you press the START key. ...

item: 713  |  http://tmi.yokogawa.com/kr/service-support/faqs/713/