The menu screen and messages are cleared when commands such as "start" and "cursor on" are issued. ...
item: 1203 | http://tmi.yokogawa.com/mx/service-support/faqs/1203/
How Much Lag Time is There When Inputting or Outting a Signal from the EXT Terminal of the WE7000 Measurement Station?
The lag time from the external terminal to the bus trigger / time base is as follows:When inputting from TRIG IN:TRIG IN -> bus trigger signal : 40 ns (typical value)TRIG IN -> time base signal : 25 ns (typical value)Whe...
item: 1139 | http://tmi.yokogawa.com/mx/service-support/faqs/1139/
Why Can't I set the DL850/DL850V to 100 ns/div?
The DL850/DL850V cannot be set to 100ns/div when the following modules are installed: 720240 CAN Bus Monitoring Module 720220 16 Channel Voltage Input Module Removing these modules permits the DL850/DL850V to be adjus...
item: 845 | http://tmi.yokogawa.com/mx/service-support/faqs/845/
Is it Possible to Bypass the Router (Subnet) and Connect to the Measuring Station with the Ethernet Module?
Yes, you can bypass the router (subnet) and connect using the WE7000 control software ver. 4 and later. However, this is limited to the WE7052 Fast Ethernet module....
item: 1157 | http://tmi.yokogawa.com/mx/service-support/faqs/1157/
How can I Improve the Signal Quality of the G3 Frequency Function on the DL850?
For the DL850 G3 Frequency Function, you will need to set the hysterisis of the edges that are used to detect the periods. This will increase the tolerance for noise and improve measurement reliability.To configure the ...
item: 1039 | http://tmi.yokogawa.com/mx/service-support/faqs/1039/
Why Can't the SL1000 Acquisition Software Detect my USB Connection?
Even though the correct USB drivers have been installed, and the device displays correctly in the Windows Device Manager Window, the SL1000 Acquisition Software fails to detect the SL1000 main unit. Solutions: Verify th...
item: 675 | http://tmi.yokogawa.com/mx/service-support/faqs/675/
Can Xviewer Combine the Binary Format Data Files Saved by Different Products?
Yes, Xviewer has the capability of opening multiple files from different DL, SL, and WE series instruments into a single window display. The only requirement for viewing files from different products is that the sample ...
item: 539 | http://tmi.yokogawa.com/mx/service-support/faqs/539/
What is the Equation Used to Convert the Voltage Axis Units From dBV to V When the Power Spectrum is Displayed on the DL1540?
If Y(V) is the converted value for X(dBV), the equation is: Y = 10(X/20) from X=20*LOG(Y) ...
item: 1191 | http://tmi.yokogawa.com/mx/service-support/faqs/1191/
Why Does the Trigger Activate Repeatedly When the Detection Level is set to "H Active" and External Trigger Input is "H" for the WE7251/WE7271 Digitizer Module?
The external trigger input (TRIG IN) as well as external input/output (EXT. I/O) trigger signal becomes the level signal. The WE7251/WE7271 digitizer module will detect the bus trigger by that level. As long as the exte...
item: 1174 | http://tmi.yokogawa.com/mx/service-support/faqs/1174/
Is There an Arbitrary Waveform Data Creation Tool for Output on the WE7281 and WE7121 Module?
Yes, there is a software called the 707751 Arbitrary Waveform Editor. You can create waveforms using the editor's standard functions or with the mouse. You can read in files created in CSV, or in Yokogawa's proprietary ...
item: 1164 | http://tmi.yokogawa.com/mx/service-support/faqs/1164/