Is There a Limit on the eSATA External HDD on the DL850?
The maximum compatible eSATA external HDD size is 1.5 TB. The external HDD recording function performs the same as the internal HDD. The actual data transfer for internal and external HDD is 3.2 MB/s. ...
item: 735 | http://tmi.yokogawa.com/my/service-support/faqs/735/
701260 Calibration Procedure Typo
701260 Calibration Procedure Typo This should be "+1V +/- 2.5mV" and "0V +/-2.5mV". The service manuals for DL750 and for SL1400 have same mistakes in writing. one attachment...
item: 1360 | http://tmi.yokogawa.com/my/service-support/faqs/1360/
DL850 G3 Power Measurement on a Sinusoid-Current Burst
The DL850 G3 Power Measurement option can be used to calculate the power of a Sinusoid-Current Burst. Please download and view image for a screenshot demonstration of the DL850 G3 Power Function.The firmware version of t...
item: 1022 | http://tmi.yokogawa.com/my/service-support/faqs/1022/
How Long Does it Take to Transfer Measured Data from the DL2700 to the PC via GPIB?
The data transfer rates for binary data is as follows: 1 kW 0.035s 10 kW 0.180s 50 kW 0.779s 100 kW 1.536s 250 kW 3.812s 500 kW 7.527s 1 MW 15.067s 2 MW 30.000s The...
item: 1227 | http://tmi.yokogawa.com/my/service-support/faqs/1227/
How do I Retrieve Data From the PZ400 Only When it is Updated on the Instrument?
You can use the "STATUS:ESSR?" command to access the extended event register and determine whether the data was updated. You can judge the data update status by referencing bit 1 (DAV) of this register. However to do so,...
item: 1239 | http://tmi.yokogawa.com/my/service-support/faqs/1239/
What is the Dead Time During Sequential Store on the DL2700?
The dead time is approximately 10µs....
item: 1221 | http://tmi.yokogawa.com/my/service-support/faqs/1221/
What is the Reference Junction Compensation Accuracy for 701265?
The specifications for Reference Junction Compensation (RJC) accuracy are listed below: Type Measurement Range Accuracy K E J T L U N -200 to 1300° C -200 to 800° C -200 to 1100° C -200 to 400° C -...
item: 1108 | http://tmi.yokogawa.com/my/service-support/faqs/1108/
Is There a Screenshot of the ScopeCorder Instrument Cart?
Please see the attached screenshot images of the ScopeCorder instrument cart. ...
item: 123 | http://tmi.yokogawa.com/my/service-support/faqs/123/
Is the Linear Averaging Function Supported by the DL1500 Series?
The averaging function supported by the DL1500 Series is exponential averaging. The models that support linear averaging (simple averaging) are the DL4000 Series, DL2700, and DL708. ...
item: 1194 | http://tmi.yokogawa.com/my/service-support/faqs/1194/
Which Bits in the Extended Event Register Corresponds to the “Check Range” LED Indicator on the WT210/WT230?
If you want to know the condition of the "Check Range" LED indicator for voltage and current, you will need to read three bits in the Extended Event Register. "Bit-6", "Bit-7", and "Bit-8" needs to be read to know the st...
item: 1095 | http://tmi.yokogawa.com/my/service-support/faqs/1095/