KNOWLEDGEBASE
(449) Results were found. Showing 71-80.
Can I use the WDF File Access API to Read *.WDF Files in MATLAB?
No, MATLAB cannot use the WDF File Access API to read *.WDF files.There is a special function in the WDF File Access API called WdfGetSacleWave64, which is used to read the data from the *.WDF file. This command uses ...
item: 909 | http://tmi.yokogawa.com/my/service-support/faqs/909/
DL850 - How to maximize repeatability of Phase Measurements
DL850 - How to maximize repeatability of Phase Measurements Use a direct-input if possible; this will remove the delay due to passive-probes 700929 being different between a pair of input channels (in phase response)...
item: 1350 | http://tmi.yokogawa.com/my/service-support/faqs/1350/
What is the Output Resolution of the D/A Analog Option for the WT100/WT200 Series?
The output resolution of the D/A option for the WT100/WT200 series is 12-bits....
item: 1232 | http://tmi.yokogawa.com/my/service-support/faqs/1232/
How do I Save DL850 Dual Capture Files in CSV Format?
The DL850 can save measured waveform data to a file in Binary, ASCII, or Floating Point format. You can use the FILE Waveform Save Menu or the SAVE MENU menu to set how files are saved.To save using the SAVE MENU method...
item: 1064 | http://tmi.yokogawa.com/my/service-support/faqs/1064/
Is There a Limit on the eSATA External HDD on the DL850?
The maximum compatible eSATA external HDD size is 1.5 TB. The external HDD recording function performs the same as the internal HDD. The actual data transfer for internal and external HDD is 3.2 MB/s. ...
item: 735 | http://tmi.yokogawa.com/my/service-support/faqs/735/
Can Xviewer decode RPM from a pulse-train?
Can Xviewer decode RPM from a pulse-train? I recommend using a Freq module for this - becuase it has many advantages; However, in a pinch, Xviewer could be used to do this: RPM = Frequency divided by pulses/rot...
item: 1362 | http://tmi.yokogawa.com/my/service-support/faqs/1362/
How do I Achieve Higher Resolution Timestamps When Viewing ASCII Files?
If you wish to view the timestamp data in high resolution, you will need to open the *.CSV file with Notepad or Wordpad. If you use Microsoft Excel to open the file, Excel will round the timestamp data, preventing you ...
item: 1110 | http://tmi.yokogawa.com/my/service-support/faqs/1110/
What is the Dead Time During Sequential Store on the DL2700?
The dead time is approximately 10µs....
item: 1221 | http://tmi.yokogawa.com/my/service-support/faqs/1221/
Does HDMI use I2C?
Yes, the HDMI pinout carries the I2C on pins 15 and 16 of the HDMI connector cable. The HDCP (content protecting encryption keys) signal is carried on pins 15 and 16. ...
item: 44 | http://tmi.yokogawa.com/my/service-support/faqs/44/
Is the Linear Averaging Function Supported by the DL1500 Series?
The averaging function supported by the DL1500 Series is exponential averaging. The models that support linear averaging (simple averaging) are the DL4000 Series, DL2700, and DL708. ...
item: 1194 | http://tmi.yokogawa.com/my/service-support/faqs/1194/