Solution:

The minimum record length which can be input is limited such that "record length x sampling interval" must be 5 milliseconds or more. When you set the record length to a value smaller than this it automatically changes to a 0 or an otherwise allowable minimum record length value. Setting the value to 0 indicates "Maximum Record Length/Number of Memory Partitions".