How to Work Around DLM2000 ASCII Record Length Limit Error Code 545

Please download the attached document for detailed instructions on how to work around the DLM2000 ASCII record length limit Error Code 545...

item: 206  |  http://tmi.yokogawa.com/nz/service-support/faqs/206/

Is There a Method to Perform Deskew on a DL850 When Using a Current Probe?

It is not necessary to perform deskew on a DL850 when using a current probe. The DL850 does have enough resolution to perform deskew and will not be able to find the time offset. Typically the deskew value ranges from ...

item: 1034  |  http://tmi.yokogawa.com/nz/service-support/faqs/1034/

Is There a Limit on the eSATA External HDD on the DL850?

The maximum compatible eSATA external HDD size is 1.5 TB. The external HDD recording function performs the same as the internal HDD. The actual data transfer for internal and external HDD is 3.2 MB/s. ...

item: 735  |  http://tmi.yokogawa.com/nz/service-support/faqs/735/

What is the Output Resolution of the D/A Analog Option for the WT100/WT200 Series?

The output resolution of the D/A option for the WT100/WT200 series is 12-bits....

item: 1232  |  http://tmi.yokogawa.com/nz/service-support/faqs/1232/

Why is the DL850 Dual Capture List Out of Order Chronologically?

On the DL850, the list starts with #0000 and corresponds to the last triggered waveform captured. The next entry, #0001, corresponds to the first triggered waveform captured and proceeds to list the captured waveforms ...

item: 1104  |  http://tmi.yokogawa.com/nz/service-support/faqs/1104/

How Long Does it Take to Transfer Measured Data from the DL2700 to the PC via GPIB?

The data transfer rates for binary data is as follows:                          1 kW 0.035s 10 kW 0.180s 50 kW 0.779s 100 kW 1.536s 250 kW 3.812s 500 kW 7.527s   1 MW 15.067s   2 MW 30.000s The...

item: 1227  |  http://tmi.yokogawa.com/nz/service-support/faqs/1227/

How do I Achieve Higher Resolution Timestamps When Viewing ASCII Files?

If you wish to view the timestamp data in high resolution, you will need to open the *.CSV file with Notepad or Wordpad. If you use Microsoft Excel to open the file, Excel will round the timestamp data, preventing you ...

item: 1110  |  http://tmi.yokogawa.com/nz/service-support/faqs/1110/

What is the Sanitation Procedure for Yokogawa Oscilloscopes?

The sanitation procedure for various Yokogawa oscilloscopes have been documented below. Please download the document or pdf that refers to your Yokogawa oscilloscope....

item: 28  |  http://tmi.yokogawa.com/nz/service-support/faqs/28/

How do I Save DL850 Dual Capture Files in CSV Format?

The DL850 can save measured waveform data to a file in Binary, ASCII, or Floating Point format. You can use the FILE Waveform Save Menu or the SAVE MENU menu to set how files are saved.To save using the SAVE MENU method...

item: 1064  |  http://tmi.yokogawa.com/nz/service-support/faqs/1064/

Can Xviewer decode RPM from a pulse-train?

Can Xviewer decode RPM from a pulse-train?   I recommend using a Freq module for this - becuase it has many advantages;   However, in a pinch, Xviewer could be used to do this:   RPM = Frequency divided by pulses/rot...

item: 1362  |  http://tmi.yokogawa.com/nz/service-support/faqs/1362/