How Do I Resolve the "LACK" Measurement Reading in DAQLOGGER When Connected to the WT230 via GateWT?
Solution:
When making a WT230 RS232C connection using GateWT, please verify the following RS232C communication settings on the instrument:
- Mode = 488.2
- Hand = 0
- For = 0
- Baud Rate = 9600
- Terminator Cr+Lf
Even though you can run GateWT and retrieve the tags from the WT230, the readings in DAQLOGGER will display "LACK" if the terminator is not set to Cr+Lf.