Can a USB to Serial Adapter/Converter be Used with a RS232C (DB25) Port on the WT210/WT230?

Although Yokogawa does not have any official recommendation for USB to Serial converters, it is possible to use some USB to RS232 DB9 Adapter Cables. If you choose to use a laptop PC without RS232C port, please use a PC...

item: 1079  |  http://tmi.yokogawa.com/ph/service-support/faqs/1079/

Power Analyzer Accuracy and Basic Uncertainty Calculator

The Power Analyzer Accuracy and Basic Uncertainty Calculator can be used to determine the uncertainty in voltage, current, and active power (watts) measurement values for various frequency ranges and wiring systems. To u...

item: 417  |  http://tmi.yokogawa.com/ph/service-support/faqs/417/

Why Doesn't WTViewer Communicate with my WT Series Power Meter via GPIB?

There are several items you will need to check and verify to solve this issue. Verify the GP-IB connectionSome instruments have a D/A output connector located next to the GP-IB connector. There have been some instance...

item: 158  |  http://tmi.yokogawa.com/ph/service-support/faqs/158/

Is it Possible to Change the Maximum Number of Harmonic Orders on the WT210/WT230?

No, it is not possible to change the maximum number of harmonic orders when calculating THD. On the WT210/WT230, the THD values are calculated based on 50 orders. To calculate THD on the WT210/WT230, there are two metho...

item: 1046  |  http://tmi.yokogawa.com/ph/service-support/faqs/1046/

Does the WT210/WT230 Conform to Energy Star's Guideline for Computers Ver. 4 approved meters?

Yes, the WT210 will meet or exceed all requirements set by the Energy Star guideline. Please see the attached documents for more details. ...

item: 73  |  http://tmi.yokogawa.com/ph/service-support/faqs/73/

How Do I Resolve the "LACK" Measurement Reading in DAQLOGGER When Connected to the WT230 via GateWT?

When making a WT230 RS232C connection using GateWT, please verify the following RS232C communication settings on the instrument: Mode = 488.2 Hand = 0 For = 0 Baud Rate = 9600 Terminator Cr+Lf Even though you can run Ga...

item: 939  |  http://tmi.yokogawa.com/ph/service-support/faqs/939/

How do I Update the Voltage and Current Range on the WT100/WT200 Series Instrument Using Pre-IEEE488.2 Commands?

To change the voltage range on element 1 to 30 V, send the "CONFIGURE:VOLTAGE:RANGE 30V" command to the measuring instrument.To change the current range on element 1 to the 1 A range, send the "CONFIGURE:CURRENT:RANGE 1A...

item: 1252  |  http://tmi.yokogawa.com/ph/service-support/faqs/1252/

What is the Output Resolution of the D/A Analog Option for the WT100/WT200 Series?

The output resolution of the D/A option for the WT100/WT200 series is 12-bits....

item: 1232  |  http://tmi.yokogawa.com/ph/service-support/faqs/1232/

Is the DA Output Terminal Connected to the Case of the WT Series Power Analyzer?

The WT3000 D/A output terminal is electrically isolated from the case. For all other models, the D/A output terminal is connected to the case. ...

item: 650  |  http://tmi.yokogawa.com/ph/service-support/faqs/650/

Why is the Peak Value and Crest Factor Unstable?

The peak value and crest factor may be unstable if they have not been captured accurately. If the peak value is not stable, neither will the crest factor be stable. The cause is the difficulty in capturing the narrow spi...

item: 1264  |  http://tmi.yokogawa.com/ph/service-support/faqs/1264/