CD (CD-ROm, CD-R, CD-RW) TA320

CD-ROM, CD-R, CD-RW

Item TA320 Advantage
Optical Disk & Pickup Evaluation
TA320 can measure and analyze all T data.
Tm Ta320 13
The data from the CD is a pulse width modulated (EFM) signal which contains the data from 3T to 11T (T being the Read Clock period.).
The pulse width jitter influences the read error.

The pulse width and jitter of all data (from 3T to 11T) for both Pit and Land are fixed by a CD Standard.

We produceTA320/520 Optical disk Intersymbol Interference Analysis software that is conform to the disk Standard.

TA320 Setting
  • Mode : Hardware histogram
  • Function : Pulse width
  • Display : Histogram
  • Multi-Window function

Example

Tm Ta320 14
With the Histogram display, It is possible to see the distribution of all data so you can determine the cause of the jitter.

Optical Disk Analysis Software

Tm Ta320 15
With this software it is possible to analyze the optical disk data for every recording length (T) which is obtained by the TA320. This gives you all the data, from 3T to 11T, in the EFM signal which is used in CD's and MD's. With this software, you can confirm the Effect Length Error and the Jitter of every T data at a glance.
This is a very useful tool to judge the disk data because this software conforms to the disk Standard for CD's and it runs on Windows.
Note: See TA320/520 Optical disk Intersymbol Interference Analysis software

The image of CD, DVD signal (EFM signal)

Tm Ta320 16
The signal of CD or DVD is as below. This image is monitored by DL1540 with pulse width trigger.
Tm Ta320 17
The image which is monitored by TA320 is as follows. Some histograms are seen as pulse width data.

Related Products & Solutions

TA320 Time Interval Analyzer

The TA320 Time Interval Analyzer is an instrument which can measure the period, pulse width and time interval of a signal continuously at a maximum rate of 14 MS/s with a maximum resolution of 100 picoseconds. The large liquid crystal display panel is equipped with a touch-sensitive screen, enabling an intuitive, smooth operation in measurement and data analysis. A high screen-update rate allows the immediate display of analysis results.