To perform a shunt calibration using the 701271 strain gage module and the ScopeCorder, please complete the following procedures:
  1. Connect the strain module 701271, bridge head 701957/701958 and the strain gauge 
  2. Attach the shunt resistor to the bridge head
  3. Set the range unit, range, excitation voltage, and gauge factor
  4. Press the Exec Balance soft key
  5. Enter the reference strain value corresponding to the attached resistor for P2:Y
  6. Select the Unit
  7. Execute the Shunt Cal to perform the calibration
  8. After the calibration is done, the P1:X and P2:X values as well as the Upper and Lower Scale will have changed
  9. If negative strain direction is selected, the P2:Y value must be negative and the P2:X value will also be negative after calibration

Please refer to the attached file for more details on shunt calibration.

Related Products & Solutions

SL1000 High-Speed Data Acquisition Unit

The SL1000 is a modular data acquisition system that delivers independent, isolated channel hardware at high speeds up to 100 MSamples per channel, with no compromise in bit resolution, memory depth, or data streaming performance. Choose from a wide variety of I/O modules for electrical and physical sensor signal measurements to meet the needs of any medium channel count application.

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.

Data Acquisition (DAQ)

Data acquisition (DAQ) is the measurement, recording, analyzing, and presentation of real world phenomena. It includes electrical measurements such as voltage, current, power as well as measurements through sensors and transducers including temperature, pressure, strain, vibration, and more.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.

Oscilloscopes

Oscilloscopes are a common type of test instrument used to capture, analyze, and troubleshoot electrical or real world physical signals. Oscilloscopes observe the change of electrical signals over time, continuously graphed on a display as voltage or amplitude vs. time.

ScopeCorders

A ScopeCorder is an instrument combining a mixed signal oscilloscope and portable data acquisition recorder into a modular platform designed to capture both high-speed transients and low speed trends. Yokogawa’s ScopeCorder product family provides flexible and high-performance multi-channel test instruments by combining a variety of signal conditioning input modules, on-board calculations, and deep data acquisition storage into a off-the-shelf data acquisition solution.

Top