What is the Dead Time During Sequential Store on the DL2700?

The dead time is approximately 10µs....

item: 1221  |  http://tmi.yokogawa.com/ru/service-support/faqs/1221/

DLM2000 SNTP Example Setup

SNTP (SNTP)The DLM2000 clock can be set using Simple Network Time Protocol (SNTP). The DLM2000 can be configured to automatically adjust its clock when it is turned on. SNTP Server (SNTP Server)Specify the IP address of...

item: 835  |  http://tmi.yokogawa.com/ru/service-support/faqs/835/

What is the Reference Junction Compensation Accuracy for 701265?

The specifications for Reference Junction Compensation (RJC) accuracy are listed below:        Type     Measurement Range Accuracy K E J T L U N -200 to 1300° C -200 to 800° C -200 to 1100° C -200 to 400° C -...

item: 1108  |  http://tmi.yokogawa.com/ru/service-support/faqs/1108/

Can I use the WDF File Access API to Read *.WDF Files in MATLAB?

No, MATLAB cannot use the WDF File Access API to read *.WDF files.There is a special function in the WDF File Access API called WdfGetSacleWave64, which is used to read the data from the *.WDF file. This command uses ...

item: 909  |  http://tmi.yokogawa.com/ru/service-support/faqs/909/

DL850 G3 Power Measurement on a Sinusoid-Current Burst

The DL850 G3 Power Measurement option can be used to calculate the power of a Sinusoid-Current Burst. Please download and view image for a screenshot demonstration of the DL850 G3 Power Function.The firmware version of t...

item: 1022  |  http://tmi.yokogawa.com/ru/service-support/faqs/1022/

How do I Retrieve Specific Items From the PZ4000 and WT1600 Using the Communications Commands?

Send the "NUMERIC:FORMAT:ASCII" command This sets the data format for the data you want to read out. Measured data read out using the "NUMERIC:NORMAL:VALUE?" command is output as an ASCII string. Send the "NUMERIC:NORM...

item: 1244  |  http://tmi.yokogawa.com/ru/service-support/faqs/1244/

Is the Linear Averaging Function Supported by the DL1500 Series?

The averaging function supported by the DL1500 Series is exponential averaging. The models that support linear averaging (simple averaging) are the DL4000 Series, DL2700, and DL708. ...

item: 1194  |  http://tmi.yokogawa.com/ru/service-support/faqs/1194/

How do I Recover From a DL850 Firmware Upgrade Failure?

If during the DL850/DL850 firmware upgrade process, the display screen turns black, blue, or something in between and the installation failed, please complete the following steps. You will need a USB printer cable to com...

item: 844  |  http://tmi.yokogawa.com/ru/service-support/faqs/844/

Which Bits in the Extended Event Register Corresponds to the “Check Range” LED Indicator on the WT210/WT230?

If you want to know the condition of the "Check Range" LED indicator for voltage and current, you will need to read three bits in the Extended Event Register. "Bit-6", "Bit-7", and "Bit-8" needs to be read to know the st...

item: 1095  |  http://tmi.yokogawa.com/ru/service-support/faqs/1095/

701260 Calibration Procedure Typo

701260 Calibration Procedure Typo   This should be "+1V +/- 2.5mV" and "0V +/-2.5mV". The service manuals for DL750 and for SL1400 have same mistakes in writing. one attachment...

item: 1360  |  http://tmi.yokogawa.com/ru/service-support/faqs/1360/