WT210/WT230 DIGITAL POWER METERS
![]() |
You may also be interested in viewing the newer release of this product. WT300 Series Digital Power Meters |

The WT210/WT230 has five times the high-frequency range and approximately twice the basic accuracy as the previous generations plus a high-speed data acquisition rate and a compact design make this digital power meter highly desirable.
What's New: The WT210 is an Accepted Device by Standard Performance Evaluation Corporation (SPEC) used for Power Efficiency Benchmarking!
For North American users please click here to learn more.
page links
Details
section links
Features
- Maximum input with assured accuracy: 26 A
- Basic accuracy: 0.1%
- DC measurement: 0.5 Hz to 100 kHz frequency range
- Compact design (half-rack size)
- 5 mA range for very low current measurements (model WT210 only)
- Line filter function
- High-speed data update (as fast as 10 readings per second)
- Harmonic measurement function available
- User calibration capability
- Large-current measurement capability using external sensor input
- Basic Characteristics
- Comparison with Former Models
- Easily Acquire and Manage Power Measurement Data from Your PC
The WT230's advanced specifications and its wide range of functions let you handle all your measurement applications from low-frequency equipment to high frequency inverters using a single power meter.
Click here to view a white paper on WT210/WT230 power meters.
Function
Wide Frequency Range
Wide Frequency Range Lets You Work on a Variety of Different Applications
- Low-frequency Equipment Low-frequency measurements starting at 0.5 Hz can be used for evaluations of cycloconverters and when a motor are started.
- Commercial Power Supplies Basic Power Accuracy: 0.1%
Power accuracy is even better than in fomer WT200/WT130 series.
- Inverters 100 kHz power frequency range
Now you can obtain more precise measurements on high-frequency equipment such as inverters.
Accuracy Is Assured Between 1% and 130%

Capture a Variety of Signal Types
| Surge Current and Maximum Load State |
This function lets you keep, on the display, voltage and current peak values, voltage and current rms values, and maximum values for active power, apparent power, and reactive power.
| Half-wave Rectification, Intermittent Control, Distortion Waves |
In addition to using DC inputs, you can obtain precise measurements of signals containing DC components, such as intermittent signals and half-wave rectification signals.

| Constantly Changing Signals |
With measurement intervals as short as 0.1 second, you can capture transient phenomena with a fine level of detail. You can also reduce the time per measurement for increased through put in production testing.
| Noisy Signals |
This function lets you measure fundamental wave rms values for inverter output voltages.
| Instead of taking notes, you can use the internal memory to store and recall settings and field measurement data. |
*3: Popular functions on the WT200 were incorporated into WT210 and WT230.
Powerful Tools for Energy Measurement
| Extended Energy Measurement Applications |
Time can be set between 1 second and 10,000 hours (416 days) in 1-second increments.
| Battery Equipment Applications |
Power and current values can be integrated separately for positive and negative polarities. Integrated values are shown with the decimal point moving according to the integrated value.
| Intermittent Control Equipment Applications |
| The power of intermittent control equipment changes significantly over time. The average active power in intermittent operations can be displayed, which is highly effective for consumed-power measurements. |
3: Popular functions on the WT200 were incorporated into WT210 and WT230
Rear Panel

One Unit Handles Standby Low-power Measurements (WT210 only)
Wide Range of 5 mA to 20 A
The built-in 5 mA range lets you measure currents as low as 25 µA. This makes it possible to measure very low currents on such things as intermittent control equipment. The wide current range (5 mA to 20 A) means a single power meter can be used for applications such as Energy Star® measurements, to measure everything from standby-power to rated-power. Benefit of Assured Accuracy from 1% of Rng

Assured accuracy from 1% allows measurement of widely-varying power consumption with a fixed range.
Comparision with Former Models
| WT200/WT130 | WT210/230 | |
| Voltage input terminal | Bindging Post | Plug-in terminal (safety terminal) |
| External input terminal | Plug-in terminal (safety terminal) | BNC |
| Voltage and current basic accuracy |
0.25% of rng | 0.2% of rng |
| Power basic accuracy | 0.3% of rng (WT200) 0.35% of rng (WT130) |
0.2% of rng |
| Frequency range | DC, 10 Hz to 20 kHz | DC, 0.5 Hz to 100 kHz |
| Assured accuracy range | 10% to 130% of range rating | 1% to 130% of range rating |
| Display updating interval | 0.25 second (fixed) | 0.1/0.25/0.5/1/2/5 seconds |
| V, A, W display digits | 4 digits (WT130) 5 digits (WT200) |
5 digits |
| Line filter function | No | Yes (fc = 500 Hz) |
| Frequency filter function | Yes (fc = 300 Hz) | Yes (fc = 500 Hz) |
| Key lock | No | Yes |
| Harmonic measurement display updating interval |
Approximately 3 seconds | 0.25/0.5/1/2/5 seconds |
| Remote signals when comparator is installed |
EXT HOLD and EXT TRIG are added. EXT START, EXT STOP, EXT RESET, and INTEG BUSY are not added. |
All six signals listed to the left are added. Pin assign is changed. |
| Online data format | ASCII | ASCII, binary |
| Waveform data communications output |
No | Yes (need /HRM) |
| Addressable mode B for GP-IB communications |
Yes | No |
| Display digits (factory default) |
4 digits | 5 digits |
| Online output data digits (factory default) |
4 digits | 5 digits |
Functions Included with the WT200 (but Not Included with the WT130) and Included with the WT210/WT230
- MAX hold function
- Moving decimal point display based on integrated power value
- 10,000-hour maximum integration time
- Integration with few data omissions
- Average active power display
Model Numbers
| Model | Description |
|---|---|
| 760401 | WT210 single-input element model |
| 760502 | WT230 2-input element model |
| 760503 | WT230 3-input element model |
Specifications
Application Notes
| Name | Description | |
|---|---|---|
| How Green is your Software? | Want to know how green is your Software? Intel® Energy Checker and Yokogawa can help you! The Intel® Energy Checker SDK provides software developers a simple way to measure the energy efficiency (greenness) of their Software by exposing the necessary and basic measurements made by Yokogawa’s highly accurate and reliable Power Analyzers. |
Learn More |
| Benefit of Assured Accuracy from 1% of Rng | Benefit of Assured Accuracy from 1% of Rng | Learn More |
Software
| Name | Description | |
|---|---|---|
| Control Module Download for GP-IB Controller Module WE7021 - For WT200 Digital Power Meter | This is an expansion module that allows operation identical to WE7000 modules by combining the WE7021 with the YOKOGAWA WT200 Digital Power Meter. See the ZIP file (WT200.pdf) for user instructions. |
y-Link |
| Control Module Download for GP-IB Controller Module WE7021 - For WT210 Digital Power Meter | This is an expansion module that allows operation identical to WE7000 modules by combining the WE7021 with the YOKOGAWA WT210 Digital Power Meter. See the ZIP file (WT210.pdf) for user instructions. |
y-Link |
| Control Module Download for GP-IB Controller Module WE7021 - For WT230 Digital Power Meter | This is an expansion module that allows operation identical to WE7000 modules by combining the WE7021 with the YOKOGAWA WT230 Digital Power Meter. See the ZIP file (WT200.pdf) for user instructions. |
y-Link |
| LabVIEW Drivers for WT200/WT210/WT230 | LabVIEW drivers for WT200/WT210/WT230 | y-Link |
| Multi-Item Display/Recording Software | The software allows you to enter the voltage and current range, and other settings in the WT series instrument's display panel. Also you can display all measurement item data (normal and calculated) on a PC, and save it to an ASCII file. |
y-Link |
| Power Consumption Measuring Software | For use with WT3000, WT1800,WT1600*1, WT500, WT210 and WT310. Conforms to IEC62301 Ed2.0(2011) and EN 50564:2011 testing methods. *1 The WT1600 discontinued sales | y-Link |
| WTViewer for WT210/WT230 (learn more) | Easily Acquire and Manage Power Measurement Data from Your PC | Learn More |
Reviews
Justin Hampton, Nordyne, LLC
We use the WT210 exclusively for single-phase inline measurements and we have found them to be solid and dependable.
However; the WT230 3-phase meter has been a lifesaver. The selectable wiring scheme is indispensable for us as most of our 3-phase equipment uses the 3-phase 3-wire scheme. This has eliminated the confusion that lead to errors in testing from the 3-phase 4-wire meters we had previously used. As far as I’m aware Yokogawa is the only company offering this option in an inline meter capable of carrying a 30A load.