Why Can't the Width level of the Window Trigger on the DL1500 Series be set to Below the Voltage Sensitivity?

Solution:

The width level of the Window trigger cannot be set to a value less than the voltage corresponding to 1 division at the specified voltage sensitivity. For example, if the voltage sensitivity is set to 500 V/div, then the width level cannot be set below 500 V. This limitation is due to the fact that the trigger sensitivity of the DL1500 Series is 1 Divp-p.

Rate This Article

Document Quality?
Answered Your Question?