AQ2200-271 OPTICAL RETURN LOSS MODULE (DISCONTINUED)
AQ2200-271 is a module for measuring optical return loss (ORL) of optical components such as optical filter and optical splitter. Insertion loss (IL) can also be measured simultaneously.
page links
Details
section links
Features
- Simultaneous measurement for both ORL and IL
- High dynamic range : 65 dB or more
- Measurement parameters can be preset (up to 9 sets)
- Built-in application function allows a simple operation
- Measurement can be automated through GP-IB, LAN, or USB
| Model | Suffix Code | Specification |
| 735185 | AQ2200-271 ORL Module | |
| -SA | Optical Fiber: SMF | |
Option
- ORL Master cord
Specifications
| Specifications | Specifications | ||
| System Configuration | Light source | AQ2200-141 or AQ2200-142 | |
| Sensor | AQ2200-211 | AQ2200-221 | |
| Wavelength Range | 1310±30 nm / 1550±30 nm | ||
| Measurement Dynamic Range | 65 dB or more¹ | 55 dB or more¹ | |
| Relative Uncertainty | RL: 50 dB or less | ±0.3 dB² | ±0.7 dB² |
| RL: 60 dB or less | ±0.6 dB² | - | |
| Applicable Fiber | SMF | ||
| Connector Type (to Light Source) | FC/PC | ||
| Connector Type (to DUT) | SC/Angled PC | ||
2: ±2σ