If your DLM2000 has firmware version 1.245 or later, then you can use a USB flash drive to upgrade the firmware of your unit. Note:DLM2000 refers to models DLM2024, DLM2022, DLM2034, DLM2032, DLM2052, DLM2054...
item: 523 | http://tmi.yokogawa.com/sg/service-support/faqs/523/
How do I Retrieve Specific Items from the WT100/WT200 Using IEEE488.2 Commands?
Send the "MEASURE:NORMAL:ITEM:PRESET:CLEAR" command to the measuring instrument. This command turns all items for output OFF. Therefore no items will be output if you send the "MEASURE:NORMAL:VALUE?" command. Send the ...
item: 1247 | http://tmi.yokogawa.com/sg/service-support/faqs/1247/
How do I set a Gage Factor Greater Than 2.2 for the 701271 Strain Input Module?
The Gage Factor setting for the 701271 Strain Input Module can be directly inputted within the range of 1.90 to 2.20. However, it is possible to indirectly set the Gage Factor to a larger value by setting the constant A ...
item: 569 | http://tmi.yokogawa.com/sg/service-support/faqs/569/
What is the Maximum Visible Light Source Distance That can be Seen by the AQ7275?
The maximum visible light source distance that can be seen by the AQ7275, through a standard SM fiber, is approximately 5 km or 3.1 miles. ...
item: 518 | http://tmi.yokogawa.com/sg/service-support/faqs/518/
How do I Read the Measurement Values From the CSV File Saved by a WT1800 Using MATLAB?
A quick and simple way of accessing the measurement data from a CSV file saved by the WT1800 is to use the MATLAB import wizard. In MATLAB, simply click on File > Import Data.... You can also use the MATLAB built-in func...
item: 1278 | http://tmi.yokogawa.com/sg/service-support/faqs/1278/
What is the Delay Time of the External Trigger Input on the DL2700?
There is zero delay because the external trigger circuit on the DL2700 is located before the A/D converter. The timing is approximately 10 ns faster than the trigger captured on the measured waveform, since there are no ...
item: 1226 | http://tmi.yokogawa.com/sg/service-support/faqs/1226/
Is There a Limitation on the Number of Characters That can be Used on the DL1540?
Up to 25 characters can be entered and displayed. If you attempt to edit a preexisting comment, only the first nine characters are displayed. In this case, you must reenter characters after the ninth character. ...
item: 1215 | http://tmi.yokogawa.com/sg/service-support/faqs/1215/
Why Can't I Make Measurements With Either WE7141 Universal Counter or WE7261/WE7262 Digital I/O Module?
With either the WE7141 100 MHz universal counter module or the WE7261/WE7262 32-bit digital I/O module, if you are performing continuous measurements with more than three of the same modules you must set the sampling int...
item: 1176 | http://tmi.yokogawa.com/sg/service-support/faqs/1176/
Can the WE7000 Main Unit be Operated on DC Power?
At present, a separate DC/AC inverter is required. In case of the WE800 measurement station (eight slots), it has a maximum power consumption of 550VA. ...
item: 1138 | http://tmi.yokogawa.com/sg/service-support/faqs/1138/
Can we upgrade the DLM2000 memory to /M2?
Can we upgrade the DLM2000 memory to /M2? No – DLM2000 Memory is factory only (must be ordered at time of new purchase); there is not a mod available for this type of upgrade. ...
item: 1310 | http://tmi.yokogawa.com/sg/service-support/faqs/1310/