Is There a Way to Recover a DL850 WDF File That Cannot be Opened in Xviewer?

If the DL850 WDF file was saved through the hard disk recording feature (HD Recording), then the file may be recovered by using the Finalize feature in Xviewer File Utility. Files that were saved through the normal save ...

item: 1284  |  http://tmi.yokogawa.com/sg/service-support/faqs/1284/

Is it Possible to Bypass the Router (Subnet) and Connect to the Measuring Station with the Ethernet Module?

Yes, you can bypass the router (subnet) and connect using the WE7000 control software ver. 4 and later. However, this is limited to the WE7052 Fast Ethernet module....

item: 1157  |  http://tmi.yokogawa.com/sg/service-support/faqs/1157/

Can I Input Tens to Hundreds of Volts to the A/D Modules for the WE7000 Series?

The WE7241 digital temperature measurement module can measure up to ±50V. The WE7271/WE7272 4-CH, 100 kS/s Isolation digitizer module up to ±35V. The WE7275 2-CH, 1 MS/s Digitizer module accepts up to ±350V....

item: 1141  |  http://tmi.yokogawa.com/sg/service-support/faqs/1141/

Why Can't I set the DL850/DL850V to 100 ns/div?

The DL850/DL850V cannot be set to 100ns/div when the following modules are installed: 720240 CAN Bus Monitoring Module 720220 16 Channel Voltage Input Module Removing these modules permits the DL850/DL850V to be adjus...

item: 845  |  http://tmi.yokogawa.com/sg/service-support/faqs/845/

How do I Monitor the Start/Stop Status of the DL716 Using Communication Commands?

It is possible to monitor the START/STOP status of the DL716 using communication commands. To do this, you will have to send the ":STATus:CONDition?" (without quotations) command to the instrument, in order to query the ...

item: 1050  |  http://tmi.yokogawa.com/sg/service-support/faqs/1050/

Are LC connectors available for the AQ7275?

The SU2005A-LCC universal connector is physically compatible with AQ7275, but we do not recommend it because we expect poor connection repeatability and return loss. We cannot guarantee the performance of AQ7275 when the...

item: 667  |  http://tmi.yokogawa.com/sg/service-support/faqs/667/

Does the Header File .HDR Saved by a ScopeCorder Contain the Instrument Settings?

Unfortunately, the header file .HDR saved by the ScopeCorder does not contain the instrument settings. If the .WVF and .HDR file pair was saved by the DL750/DL750P/SL1400, you may reload the WVF file back into the instru...

item: 315  |  http://tmi.yokogawa.com/sg/service-support/faqs/315/

What is the Update Rate of the Waveform Parameters for the DL1540?

The update rate is 1 to 2 Hz when all measurement items are specified on Trace1 through Trace4. The update rate is 5 to 6 Hz when only a few measurement items are specified on Trace1. ...

item: 1207  |  http://tmi.yokogawa.com/sg/service-support/faqs/1207/

How Can I Programatically Verify if the Calculations for Waveform Parameters are Completed?

There are two methods to programmatically determine if the calculations for waveform parameters is completed on the DL series oscilloscopes. Send the :MEAS:WAIT? command. This will ensure that no other commands are exec...

item: 1041  |  http://tmi.yokogawa.com/sg/service-support/faqs/1041/

What is the Limit for Each Module That can be Placed in a DL850?

The number of each module that can be placed in a DL850 is limited to a maximum of: Voltage Scanner Module: 8 CAN Bus Monitoring Module: 2 (must be installed in slot 7 & 8) Logic Input Module: 8 100 MS/sec Module: 4 (mu...

item: 709  |  http://tmi.yokogawa.com/sg/service-support/faqs/709/