How do I Calculate Channel to Channel Skew on the DLM2000 Series Oscilloscope?

The rated specification for channel to channel skew on the DLM2000 is 1 ns or less, where 1 ns typically denotes the worst case scenario.

To measure channel to channel skew on the DLM2000 series oscilloscope, you will need to verify the test conditions for both channels are the same. This means using the same probe type, the same input device, and sampling both channels at the same rate.

For a simple method on how to measure channel to channel skew, follow the procedures listed below:
  1. Use a Function Generator as the input device and output a 1MHz, 500mVpp, Pulse wave
  2. Connect the output of the Function Generator to a Power Divider
  3. Connect the output of the Power Divider to CH1 and CH2 on the DLM2000 (Setup should now look the same as the attached JPEG)
  4. Set the DLM2000 to 100 mV/div, Sampling Mode Repetitive, Average Mode, Count 16
  5. Set the Trigger Edge to Positive.
  6. Measure the time difference between the rising edges of CH1 and CH2

Related Products & Solutions

DLM2000 Mixed Signal Oscilloscopes

200, 350, and 500MHz mixed-signal oscilloscopes for every engineer. Best-in-class performance in usability, acquisition, analysis, and display—all at a price you can digest. Options include serial bus, vehicle bus, and power supply analysis functions.

Mixed Signal Oscilloscopes

A mixed signal oscilloscope is an instrument enabling the simultaneous, time correlated observations and analysis of analog with digital (logic) signals, to establish causal relationships between the various areas of a device's electronics. They are purpose-built tools for troubleshooting electrical anomalies, measuring parametric values, and monitoring cause and effect relationships between signals.

Oscilloscopes

Oscilloscopes are a common type of test instrument used to capture, analyze, and troubleshoot electrical or real world physical signals. Oscilloscopes observe the change of electrical signals over time, continuously graphed on a display as voltage or amplitude vs. time.

Top