What is the Update Cycle of the D/A Output of the WE7141 100Mhz Universal Counter Module?

The update cycle changes depending on the acquisition mode and the gate time during frequency measurement. When the acquisition mode is One Shot, the update cycle is approximately 0.5-1 seconds. However for frequency measurement, it's about 10 seconds when the gate time is 10 seconds. When the method is Free Run, the update cycle occurs at each sampling interval. However for frequency measurement, it occurs approximately at each (gate time + sampling interval).

Related Products & Solutions

WE7000 PC-Based Measurement Instruments

One system, multiple instruments:  WE7000 satisfies demands for fast, reliable and precise data acquisition which uses a standard laptop or PC as its user interface. Input modules plug into an expandable measuring station. Via the optical interface, it is possible to galvanically isolate the laptop/PC.

Data Acquisition (DAQ)

Data acquisition (DAQ) is the measurement, recording, analyzing, and presentation of real world phenomena. It includes electrical measurements such as voltage, current, power as well as measurements through sensors and transducers including temperature, pressure, strain, vibration, and more.

High Speed Data Acquisition

Yokogawa high speed data acquisition systems deliver industry leading isolation, bit resolution, sampling rate, and memory depth, with independent channel hardware and easy to use software.