Xviewer 1.70 Math Issue

Xviewer 1.70 Math Issue Normally, for the math expression 'Mn' (when n...

item: 1342  |  http://tmi.yokogawa.com/th/service-support/faqs/1342/

Why Doesn't the DL1540 Recognize the Isolated Input Unit?

To use the isolated input unit, the firmware version of the DL1540 must be 1.10 or higher. Please check the version number. If the version number is lower than 1.10, the firmware must be upgraded. ...

item: 1214  |  http://tmi.yokogawa.com/th/service-support/faqs/1214/

Is it Possible to Make Continuous Measurements on the WE7000 Series With Zero Dead Time and Send the Data to PC Using the A/D Modules?

You can perform continuous measurement using the digitizer module's Free Run mode. However, data transmission speeds are limited. Please consult your sales representative for details. They can tell you whether data trans...

item: 1143  |  http://tmi.yokogawa.com/th/service-support/faqs/1143/

What is the Difference Between Offset and Position on the DL Series Oscilloscopes?

The Offset function is used to set the vertical position of a waveform and is defined by a voltage value (V). The Position function is used to set the horizontal location of a waveform and is defined by a division value ...

item: 588  |  http://tmi.yokogawa.com/th/service-support/faqs/588/

What is the Maximum Input Power for Pulsed Light Measurements on the AQ6370B?

The maximum input power that the AQ6370B can measure for pulsed light measurements is +20dBm per resolution at full span, per channel. Keep in mind that this limit is applicable to an averaged power of pulsed light, not ...

item: 161  |  http://tmi.yokogawa.com/th/service-support/faqs/161/

Does the DLM2000 Decode and Analyze J1939?

Yes, the DLM2000 can decode and analyze J1939....

item: 1305  |  http://tmi.yokogawa.com/th/service-support/faqs/1305/

INTEG1 vs INTEG2 - as demonstrated on DL850 G3

INTEG1 vs INTEG2 - as demonstrated on DL850 G3   Examples of two different ways to INTEGRATE on the DL850/G3 DSP.   One attachment. ...

item: 1311  |  http://tmi.yokogawa.com/th/service-support/faqs/1311/

Why Does the Measurement of Inverter Efficiency Exceed 100%?

This could be caused by any of the following: Is there any crosstalk (especially around the 2nd order)? Are there any effects of CMRR (especially around the 2nd order)? Is the location of measurement immediately next ...

item: 1266  |  http://tmi.yokogawa.com/th/service-support/faqs/1266/

What is the Delay Time of the External Trigger Input on the DL2700?

There is zero delay because the external trigger circuit on the DL2700 is located before the A/D converter. The timing is approximately 10 ns faster than the trigger captured on the measured waveform, since there are no ...

item: 1226  |  http://tmi.yokogawa.com/th/service-support/faqs/1226/

Can the Waveform Data Saved in .WVF Format on the DL1540 be Loaded Onto the DL1540C/DL1540CL?

It is currently not possible to load .WVF waveform data saved by a DL1540 onto the DL1540C/DL1540CL or vice versa. This is because the internal circuit is not common between the two models. It is however, possible to loa...

item: 1187  |  http://tmi.yokogawa.com/th/service-support/faqs/1187/