AQ2200-601 10 GBIT/S BERT MODULE (DISCONTINUED)
The AQ2200-601 10 Gbit/s BERT module is a compact and powerful. It offers high quality features such as built-in internal Signal Generator (SG) and Clock Data Recovery (CDR), EO/OE interfaces, and several variable signal output functions. The AQ2200 series MultiApplication Test System is designed for universal purposes; therefore various modules for various applications can be accommodated. (i.e. Optical Power Meter, Light Source and Optical Attenuator modules).
- PPG, ED and SG integrated design
- Handles multiple bit rates (9.95 Gbit/s to 11.32 Gbit/s)
- Adjust output amplitude, offset and crosspoint
- Clock and Data Recovery (CDR) function included
- Compact and lightweight
- Combine with other optical modules to create a variety of tests
- Combine with an optical modulator, receiver, and attenuator to measure error rate curves
- 64 Mbit long program pattern generation capability
- Output level range: 0.5 to 2.0 Vp-p
- Offset range: -2.0 V to +3.0 V
- Cross-point adjustment range: 30 to 70%
- Input threshold level adjustment
- Optical modulator and receiver (Optional)
page links
Details
section links
Features
- Optimize the use of lab bench space
- Take measurements in close proximity to the system/device being measured
- Easily portable
- PPG function
- Support for multiple bit rates:
9.95 Gbit/s to 11.32 Gbit/s, variable in steps of 1 kHz - Output amplitude:
0.5 Vp-p to 2.0 Vp-p, variable in steps of 10 mV - Offset:
-2.0 V to +3 V, variable in steps of 10 mV - Crosspoint:
30% to 70%, variable in steps of 1% - Data and data bar:
2 outputs - Trigger output for oscilloscope synchronization
- Port for external 10 G clock synchronization input and 1/16 or 1/64 synchronization input
It is possible to perform an error rate test with jitter deliberately added to the data pattern by inputting a synchronization signal with added jitter to this port.
- Support for multiple bit rates:
- Bit error rate test (BERT)
A BERT is a type of code error measurement made through digital communication using a PPG as the signal source and an ED as the error detector.
- ED function
- CDR function included as standard feature
The CDR signal extracts the clock from the data signal and performs synchronization when a data signal is the only input. In optical communications, the CDR function is a useful tool because the data signal is often transmitted alone. - In addition to the CDR input port, there are ports for inputting separate data and clock signals. When these ports are used, the phases of the data and clock signals must be matched externally.
- Input ranges:
0.1 Vp-p to 0.7 Vp-p (with CDR), 0.1 Vp-p to 0.6 Vp-p (without CDR)
- CDR function included as standard feature
- Test patterns
- There are two types of test patterns: pseudo-random (PRBS) patterns and program patterns.
- The PRBS pattern is selected from settings ranging from PRBS7 to PRBS31.
- In the standard configuration, the AQ2200-601 program pattern consists of 256 bits. A 64 Mbit option is also available.
- Supported bit rates
Network name Standard Standard bit rate FEC SDH/SONET STM-64/OC-192 9.953280 Gbit/s 10.664228 Gbit/s STM-16/OC-48 10.796778 Gbit/s 10G Ethernet 10 GBASE-R 10.312500 Gbit/s 11.095728 Gbit/s Fibre Channel 10 G-FC 10.518750 Gbit/s 11.317642 Gbit/s OTN OTU2 10.709225 Gbit/s
Mainframe Screenshot

1-1 Elapsed time display
- Displays the elapsed time
- Displays an indicator which indicates the logging execution status
- System clock error: Displays the system clock status
- Err on: Displays the PPG error addition ON/OFF status
- Output Displays the PPG output signal ON/OFF status
- CDR ulk:Displays the ED CDR function operation status
- Sync-los: Displays the ED synchronization status
- Bit-err: Displays the ED bit error detection status
- : Syn-los: Displays the ED synchronization loss time (in microseconds)
- Err-cnt: Displays the ED received data error count (in bits)
- Err-rate: Displays the ED received data error rate as a numerical value
- Starts measurement
- Stop mes: Stops measurement
- Error add: Turns error addition ON/OFF
- Signal output: Turns PPG data output and clock output ON/OFF
- Information: Displays the firmware version and other information
- Displays the settings. Use the tab key to switch the display
Example Applications
Optical device characteristics evaluation test

- Rise and fall times: Tr/Tf, Crosspoint position
- Overshoot percentage
- Jitter measurements: Peak to peak and RMS
- Extinction ratio
- Pulse mask test
- Output power (peak, average), light wavelength

- Jitter test
- Optical reception sensitivity test
- Optical output eye mask test
Specifications
PPG and ED| Item | Product specifications | |
|---|---|---|
| Internal clock | Internal frequency variation | Variable frequency range: 9.95 GHz to 11.32 GHz |
| Minimum setting increment: 1 kHz | ||
| External reference | Frequency | 1/64 or 1/16 bit rate |
| Input level | 0.4 to 1.0 Vp-p: 50 Ω AC | |
| External clock input | Frequency | 1/1 bit rate |
| Input level | 0.4 to 1.0 Vp-p: 50 Ω AC | |
| Trigger output | Clock trigger | 1/64 or 1/16 clock output frequency |
| Pattern triger | PRBS, PROGRAM | |
| Output level | 0.6 Vp-p ±0.3 V | |
PPG only
| Item | Product specifications | |
|---|---|---|
| Data output | Bit rate | 9.95 Gbit/s to 11.32 Gbit/s |
| Output level | 0.50 to 2.00 Vp-p (in steps of 10 mV) | |
| Offset voltage | -2 V to +3 V (in steps of 10 mV) | |
| Crosspoint variation | 30% to 70% (in steps of 1%) | |
| Connector | 3.5 mm female 50 Ω AC termination or DC termination |
|
| Clock output | output level | 0.6 Vp-p (AC coupled) |
| Offset | -2 V to +3 V (in steps of 10 mV) | |
| Connector | SMA female 50 Ω AC termination or DC termination |
|
| Transmit patterns | PRBS | 7, 9, 10, 11, 15, 23 or 31 |
| PROGRAM | (1) 16 to 256 bits (in steps of 1 bit) (2) Option for 256 - 67, 108, 864 bits (in steps of 128 bits) |
|
ED only
| Item | Product specifications | |
|---|---|---|
| DATA IN (CDR) | Bit rate | 9.95 Gbit/s to 11.32 Gbit/s |
| Input level range | 0.1 to 0.7 Vp-p | |
| Input threshold value variation | ±0.35 V (in steps of 1 mV) | |
| Connector | 3.5 mm female: 50 Ω AC coupling | |
| DATA IN | Bit rate | 9.95 Gbit/s to 10.71 Gbit/s |
| Input level range | 0.1 to 0.6 Vp-p | |
| Input value variation | ±0.3 V (in steps of 1mV) | |
| Connector | 3.5 mm female: 50 Ω AC coupling | |
| Clock input | Frequency | Frequency with same bit rate synchronized to data input |
| Input level | 0.2 to 0.6 Vp-p | |
| Connector | SMA female: 50 Ω AC coupling | |
| Receiving patterns | Same as PRBS and PROGRAM sending patterns | |
Specifications
| Name | Description | File Type | |
|---|---|---|---|
| Bulletin AQ2200-20EN | Ideal Measurement Solution for Optical Devices and Optical Transmission Systems | 1.78 MB |
Download |