Can I Convert AQ6370 Waveform Data from a Binary to a ASCII .CSV File?

It is possible to convert an AQ6370 Binary file to ASCII .CSV file using the AQ6370 Viewer Software. Please keep in mind there are different AQ637X series OSA Viewer software will load specific binary file extensions.  ...

item: 1129  |  http://tmi.yokogawa.com/tw/service-support/faqs/1129/

What is the Chopped Light Mode?

The Chopped Light mode is used: To increase measurement level sensitivity by cutting the high frequency noise To detect only the light that the LS emitted, effective for free space measurement. In free space measurement...

item: 297  |  http://tmi.yokogawa.com/tw/service-support/faqs/297/

On the AQ637X Series OSA Why is the Actual Resolution Not the Set Value?

The resolution is controlled by 3 factors: Input Slit Detraction Grating  Output Slit The resolution setting controls the output slit disk. This disk has a fixed number of slits in it. Since the setting only control on...

item: 1057  |  http://tmi.yokogawa.com/tw/service-support/faqs/1057/

What is the Largest Core Diameter Fiber I can use in the AQ6373, AQ6375 and AQ6370 series OSAs?

For the AQ6373 you can use a maximum 800 micron core fiber that must be terminated in a standard FC connector.   For the AQ6375 and AQ6370 series OSA you can use a maximum 200 micron core fiber which, again must be term...

item: 1112  |  http://tmi.yokogawa.com/tw/service-support/faqs/1112/

Is the RMS Method for Spectrum Width Calculation on the AQ6370C FOTP-127 Compliant?

The AQ6370C OSA was developed in accordance with the IEC standard: WG4 (IEC 61280-1-3) Test procedures for general communication subsystems -Central wavelength and spectral width measurement. We cannot guarantee the ...

item: 1308  |  http://tmi.yokogawa.com/tw/service-support/faqs/1308/

How do I Make High Bit Rate Modulated Signal OSNR Measurements?

Please download the attached application note for instructions on how to make High Bit Rate Modulated OSNR Measurement for the AQ6319 and AQ6370 series OSAs....

item: 298  |  http://tmi.yokogawa.com/tw/service-support/faqs/298/

How do I Resolve the Warning Message:142 WL Calibration Failed?

If you are performing wavelength calibration using the instrument's built-in light source, the warning message is an indication that the monochromator needs to be repaired. Please visit our Contact site to locate and c...

item: 401  |  http://tmi.yokogawa.com/tw/service-support/faqs/401/

Is the AQ6370 Viewer Software Compatible with Windows 7?

Yes, the AQ6370 Viewer Software is compatible with Windows 7 32-bit and 64-bit. Note: Please make sure the correct USB driver is installed for either Windows 7 32-bit or 64-bit....

item: 991  |  http://tmi.yokogawa.com/tw/service-support/faqs/991/

Why Does my AQ637X OSA Freeze Every so Often?

The AQ637X OSA has a feature called Auto Offset that it performs every 10 minutes. What this feature does is zero out the OP amps in the OSA to ensure constant level accuracy. This feature will however, stop the OSA from...

item: 1081  |  http://tmi.yokogawa.com/tw/service-support/faqs/1081/

What is the Sanitation Procedure for the AQ637X Series OSAs?

You can restore all settings for the AQ637X series OSAs to their factor defaults by performing a parameter initialization. This will clear all current parameter setting values and data as well as alignment adjustment val...

item: 224  |  http://tmi.yokogawa.com/tw/service-support/faqs/224/