How do I Read the Measurement Values From the CSV File Saved by a WT1800 Using MATLAB?
A quick and simple way of accessing the measurement data from a CSV file saved by the WT1800 is to use the MATLAB import wizard. In MATLAB, simply click on File > Import Data.... You can also use the MATLAB built-in func...
item: 1278 | http://tmi.yokogawa.com/tw/service-support/faqs/1278/
Will all DL750 Modules Work on the DL850 or SL1000?
In most cases, all the DL750 modules will be compatible with the DL850 and SL1000. However, there are some module which can be used with DL750 but cannot be used with DL850. These are typically the early DL750 modules t...
item: 841 | http://tmi.yokogawa.com/tw/service-support/faqs/841/
Can I Upgrade the DLM2000 Firmware Using USB Flash Drive?
If your DLM2000 has firmware version 1.245 or later, then you can use a USB flash drive to upgrade the firmware of your unit. Note:DLM2000 refers to models DLM2024, DLM2022, DLM2034, DLM2032, DLM2052, DLM2054...
item: 523 | http://tmi.yokogawa.com/tw/service-support/faqs/523/
How do I Upgrade the Firmware for the AQ2201/AQ2202 Frame Controller?
Please download and refer to the attached document for complete instructions on how to upgrade the firmware for the AQ2201/AQ2202 Frame Controller. ...
item: 300 | http://tmi.yokogawa.com/tw/service-support/faqs/300/
What is the Sanitation Procedure for the AQ637X Series OSAs?
You can restore all settings for the AQ637X series OSAs to their factor defaults by performing a parameter initialization. This will clear all current parameter setting values and data as well as alignment adjustment val...
item: 224 | http://tmi.yokogawa.com/tw/service-support/faqs/224/
Why Does the Measurement of Inverter Efficiency Exceed 100%?
This could be caused by any of the following: Is there any crosstalk (especially around the 2nd order)? Are there any effects of CMRR (especially around the 2nd order)? Is the location of measurement immediately next ...
item: 1266 | http://tmi.yokogawa.com/tw/service-support/faqs/1266/
How do I set a Gage Factor Greater Than 2.2 for the 701271 Strain Input Module?
The Gage Factor setting for the 701271 Strain Input Module can be directly inputted within the range of 1.90 to 2.20. However, it is possible to indirectly set the Gage Factor to a larger value by setting the constant A ...
item: 569 | http://tmi.yokogawa.com/tw/service-support/faqs/569/
How Much Residual Noise Should There be on a Yokogawa Oscilloscope?
The true noise floor of a Yokogawa oscilloscope is ≤ 2-3mVpp. The DL9000, for example, has approximately 280uV residual noise at 2mV/Div with input termination set to 50Ω. The residual noise may be measured by making a ...
item: 531 | http://tmi.yokogawa.com/tw/service-support/faqs/531/
Why are the Input Terminals on the Rear Panel of the Power Meter?
The input terminals on all Yokogawa power meteras are located on the rear panel. This takes into account safety when handling the measuring instrument. The signal input to the power meter normally carries high voltage...
item: 1269 | http://tmi.yokogawa.com/tw/service-support/faqs/1269/
What is the Step Response of the 701280 Frequency Input Module for a Sine Wave Filtered with an IIR LP Filter?
For the 701280 Frequency Input Module with a 4th order IIR 6Hz lowpass filter, a 60Hz sine wave (FM modulated with a square wave) was measured with deviations of ± 3Hz. Please download and view the attached image for a ...
item: 986 | http://tmi.yokogawa.com/tw/service-support/faqs/986/