Can Multiple GS Units be Connected in Series/Parallel to Boost the Output Voltage/Current?

It is possible to connect multiple GS units in series/parallel connections to boost the maximum voltage/current, provded the values are within the range of maximum voltage/current of those limiters. To amplify the output voltage of more than 1 unit (or 1 Ch), connect the units in series. To amplify the output current of more than 1 unit (or 1 Ch), connect the units in parallel.
 
Please refer to attached file. It is written as a case of 2 units but it is common for the case of 3 units. It is also common for the case of using CH1 and CH2 on GS820.

Related Products & Solutions

GS200 DC Voltage / Current Source

The GS200 is a low voltage/current DC source with high accuracy, high stability, and high resolution.

 

 

GS610 Source Measure Unit


The GS610 is made up of a programmable current and voltage source, a voltmeter, and an ammeter. Each function can be combined into numerous operation modes.

GS820 Multi Channel Source Measure Unit

The GS820 is a highly accurate and highly functional 2-channel programmable DC voltage/current source that incorporates voltage/current generation and measurement function.

Signal Generators, Sources and Supplies

For general purpose standalone applications or as core components in a high speed test and measurement system, Yokogawa sources and signal generators are highly accurate and functional.

Source Measure Units

A Source Measurement Unit (SMU) is a flexible test instrument which is capable of both sourcing and measuring voltage and/or current at the same time.

Precision Making

Top