700924 100 MHZ DIFFERENTIAL PROBE

The 700924 differential probe is a differential probe with a bandwidth of 100M Hz and a 100:1 or 1000:1 attenuation ratio.
page links
Details
section links
Features
Applicable for DL750/DL750P & SL1400
With increases in the speed of measured signals in recent years, it has become necessary to make floating measurements over a wide frequency band. The 700924 differential probe lets you make wide-band (100 MHz) differential input measurements by just connecting the probe to the input of a single-end input digital oscilloscope.
Oscilloscope input amplifiers are normally single-ended. They do not provide accurate measurements when used in combination with a 10:1 passive probe for measuring the gate-to-source voltage of a switching power supply, or potential difference with a common mode voltage in elements such as Insulated Gate Bipolar Transistors (IGBTs).
Furthermore, there is a risk that current will enter through the oscilloscope ground line and damage the oscilloscope.


Specifications
| Specifications |
|
| Input type | Balanced differential input |
| Frequency band |
DC up to 100 MHz (-3 dB) |
| Input attenuation ratio |
Switched between 1/100 and 1/1000 |
| Input impedance |
Approximately 4 MΩ, approximately 10 pF parallel |
| Gain accuracy |
±2% (common mode voltage: 400 V or less) ±3% (common mode voltage: 1000 V or less) |
| Maximum allowed differential voltage |
±350 V (DC + ACpeak) or 250 Vrms (1/100 range) ±1400 V (DC + ACpeak) or 1000 Vrms (1/1000 range) |
| Maximum common mode input voltage |
±1400 V (DC + ACpeak) or 1000 Vrms (both 1/100 and 1/1000 ranges) |
| Maximum input voltage |
±1400 V (DC + ACpeak) or 1000 Vrms (1/1000 range) |
| Common Mode Rejection Ratio (CMRR) |
-80 dB (60 Hz), -50 dB (1 MHz) |
| Output voltage |
±3.5 V |
| Output impedance |
Used with 1 MΩ input impedance |
| Power supplies |
Dry cells, four R6P (SUM-3); when using the AC/DC adapter, use an output voltage of 6 V/200 mA or greater/center plus |
| Dry cell life |
Approximately 2 hours under continuous use |
| External dimensions (approximate) |
207 mm x 83 mm x 38 mm¹ |
| Total length |
Approximately 1.5 meters |
| Weight |
Approximately 800 grams (excluding dry cells) |
| Approved standards |
EMC: EN55011: 1991 + A1: 1997 + A2: 1996 EN50082-1: 1992 Safety: EN61010-1: 1993 + A2: 1995 EN61010-2-031: 1994 |
Note: Dry cells and AC/DC adapter not included.
Model and Code
| Model Name | Code | Description |
| Differential probe |
700924 |
DC to 100 MHz band |
Specifications
| Name | Description | File Type | |
|---|---|---|---|
| Bulletin 7009-24E | 700924 100 MHz band differentital probe | 235 KB |
Download |