Frequency to Voltage Conversion
Frequency to Voltage Conversion Problem: A transducer produces a sine-wave output dependant on the voltage-input; here is how to decode that output with a scope or a ScopeCorder. For 3kV input, the transducer outpu...
item: 1409 | http://tmi.yokogawa.com/us/service-support/faqs/1409/
Why do I Need to use a Hub When Communicating With the Instrument via Ethernet?
The AC Power Input in all Yokogawa instruments is designed as a 3-pin connection (one of which is a GND pin). In some parts of the world, PCs are sold with AC power cables that are 2-pin. Often times this means the PC d...
item: 906 | http://tmi.yokogawa.com/us/service-support/faqs/906/
Where can I Find Documentation Describing the *.WDF File Format?
Unfortunately, we do not have an instruction manual describing the*.WDF file format structure like we did for the *.WVF files. However, we do provide a list of support options which will contain more details and explana...
item: 1042 | http://tmi.yokogawa.com/us/service-support/faqs/1042/
What is Hi-Resolution Mode?
High Resolution mode can be used to remove high frequency noise and increase vertical resolution. It achieves this by increasing the number of effective bits per data to 12-bits through digital and bandwidth filters. Th...
item: 1076 | http://tmi.yokogawa.com/us/service-support/faqs/1076/
What is Decimation and How do I use Decimation in Xviewer?
Decimation is a technique used to reduce the total number of samples. You can use Xviewer to perform decimation on your waveform data files that have the WDF/WVF/ASCII CSV extension format. Decimation reduces the total n...
item: 1029 | http://tmi.yokogawa.com/us/service-support/faqs/1029/
How Can I Effectively Achieve a Sample Rate Less Than 5 Sa/sec on the DL850?
The lowest possible device setting for sample rate on the DL850 is 5 Sa/sec. There are however, different methods that will allow you to achieve much lower sample rates which includes the use of an external clock, decim...
item: 1003 | http://tmi.yokogawa.com/us/service-support/faqs/1003/
What is the Measured Propagation Delay for the PB500 500 MHz Passive Probe (701943)?
The measured propagation delay for the PB500 500 MHz Passive Probe (701943) is approximately 8.2 ns. A simple way of measuring the propagation delay, using the DL9000, is to measure the delay between two channels usin...
item: 1109 | http://tmi.yokogawa.com/us/service-support/faqs/1109/
Is VXI-11 Required for a DLM2000 to Xviewer Connection Over Ethernet?
Yes, VXI-11 is required and needs to be enabled for the DLM2000 to connect properly to Xviewer when using an ethernet connection. If the VXI-11 option is unchecked, a connection failed message will appear. Related Inf...
item: 208 | http://tmi.yokogawa.com/us/service-support/faqs/208/
How Many Math Channels are Available on the DL6000 and DLM6000?
The DL/DLM6000 series supports up to 4 Math Channels, which can be accessed by pressing one of the M1-M4 buttons....
item: 462 | http://tmi.yokogawa.com/us/service-support/faqs/462/
What is the Sanitation Procedure for Yokogawa Oscilloscopes?
The sanitation procedure for various Yokogawa oscilloscopes have been documented below. Please download the document or pdf that refers to your Yokogawa oscilloscope....
item: 28 | http://tmi.yokogawa.com/us/service-support/faqs/28/